CONCEPTUAL · DRAFT v0.3 · NOT FOR CLINICAL USE. An educational thought experiment about how a care robot could follow an auditable protocol. It is not a medical procedure, has not been validated, and must not be used to care for anyone. Thresholds marked <set by …> are deliberately left for clinical leads and validation to set. Disclaimer
Robotic Medicine · Protocol library

RMP-PREV-001 Medical kit and equipment readiness checks: labels, seals, expiry and self-test status

AI-assisted Reviewed (AI-assisted) AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 7 review, 14 required corrections and 17 optional suggestions); all required corrections are applied in this version (see Corrections). No named human clinician has reviewed it. Not for clinical use.

Fixed rules (same in every protocol)

1 · Identity and scope

On the CMO's schedule, or after any kit has been used, the robot goes to each kit and device on the CMO-approved list in its stowage location, reads the kit ID label, seal status and printed expiry dates, and the device self-test and battery indicators as displayed, compares them with the inventory record, and sends a report to the CMO, the flight surgeon and the pharmacy lead. It never opens a kit, so it reports "LABEL, SEAL AND EXPIRY MATCH THE RECORD. Contents not checked." at best. A kit shown as "opened – not restocked" stays flagged until a human restocks and reseals it. The crew do the open-kit checks, device tests, restocking and disposal.

Why this protocol was chosen. Emergency kits only help if they are where the crew expect, sealed and in date. NASA technical brief OCHMO-MTB-012 notes that missions of 1-5 years will outlast the 1-2 year expiration dates of many medicines, and links medicine storage and inventory tracking to NASA-STD-3001 V2 7059. Volume 2 also requires that the stowage identification system works with inventory management (V2 7058), that items have dedicated locations (V2 7052), and that mission-critical items are the easiest to reach (V2 7055). A robot that reads labels and seals in place, without moving or opening anything, can take routine checking off the crew while leaving every judgment to them. Open question for review: which kits and devices are on the check list, how often, and which faults count as emergency-kit faults.

Autonomy: Max rung 2 for supervised reading of kit labels, seals, printed expiry dates and device status displays in place. The robot never opens a kit, never runs a device test, never removes or discards an item, and never calls a kit ready.

Covers

  • Scheduled checks and checks after any kit use (including after an emergency fetch, once the RMP-PHARM-001 reconciliation is done)
  • Medical kits, the defibrillator, oxygen, airway, eyewash, trauma and anaphylaxis kits, monitors and the ultrasound unit, as the CMO lists them
  • Routing to RMP-PHARM-001 for anything involving a medicine

Locked out (never done by the robot under this protocol)

  • Diagnosis or naming a cause
  • Touching the crewmember
  • Opening a kit or breaking a seal
  • Running, charging or pressing any device test or control
  • Removing, discarding or moving an item, except returning a kit to its own location on the CMO's command
  • Calling a kit or device ready, complete or verified
  • Handing over any medicine outside RMP-PHARM-001, or after a MISMATCH or CANNOT VERIFY (the item stays quarantined; only the CMO can ask for it, through the RMP-PHARM-001 logged two-step override)

Related: RMP-PHARM-001 · RMP-CARD-001 · RMP-RESP-001 · RMP-RESP-003 · RMP-INF-002 · RMP-EGR-001

2 · Roles

Medical authority
Flight surgeon (ground). Holds medical authority for every run.
Onboard lead
Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when there is no link.
Others
  • Second trained CMO: receives every alert and takes over if the CMO is the patient
  • Pharmacy lead (ground): receives every medicine and kit flag
  • Other crew: help as the CMO directs
Commercial flights
On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001 requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for spaceflight participants.) [7]

Robot may

  • Read kit ID labels, seals and printed expiry dates in place
  • Read device self-test and battery indicators as displayed
  • Compare readings with the inventory record
  • Send the report to the CMO, flight surgeon and pharmacy lead
  • Return a kit to its own location on the CMO's command

Robot may not

  • Open a kit or break a seal
  • Run or press any device test
  • Remove, discard or move items
  • Call anything ready or verified
  • Start without a human command or an accepted schedule prompt

3 · Preconditions and consent

Consent
No patient is involved; the CMO approves the check list.
Emergency consent
Any medical need stops the check at once and leaves every kit where the crew can reach it.
Withdrawal
The CMO can stop the check; the report so far stands.

4 · Required sensor inputs

IDInputSourceRequiredIf missing
IN-1Kit ID label, seal status and printed expiry dates, read in placeRobot camera / barcode readerYesLog "CANNOT READ" for that item; never assume a match.
IN-2Device self-test and battery indicators as displayedDevice displayYesLog "CANNOT READ".
IN-3Inventory record for each itemInventory management systemYesLog "no record"; flag.
IN-4The CMO's acknowledgement of the reportCMOYesLog "no response"; the report is repeated to the second trained CMO.
IN-5Link status and delayCommunications systemYesTreat as link down; the report is queued for the ground.

5–6 · Steps and decision points

Each step names who acts. Decision points sit after the step they follow and say who decides: Robot Crew Ground. Any stop rule in section 7 can interrupt any step.

  1. S1RobotRung 2

    Tell the flight surgeon and the CMO that a readiness check has started (logged and sent with the next package; URGENT only if a red flag is stated). Every URGENT alert goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.

    Done when: Notice logged and queued.

    V2 10174V1 6003V1 3006

  2. S2RobotRung 2

    Go to each kit and device on the CMO-approved list in its stowage location and, without opening or moving it, read the kit ID label, seal status and printed expiry dates, and the device self-test and battery indicators as displayed.

    Done when: Each item read, or "CANNOT READ" logged.

    V2 7058V2 7052

  3. S3RobotRung 2

    Compare each reading with the inventory record and show the result per item: "LABEL, SEAL AND EXPIRY MATCH THE RECORD. Contents not checked.", or MISMATCH with the reason (seal not intact, past printed expiry, wrong location, missing, failed self-test as displayed, low battery as displayed, no record), or CANNOT READ. A kit shown as "opened – not restocked" stays flagged until a human restocks and reseals it.

    Done when: Result shown for every item.

    V2 7059

    D1DecisionRobot decides

    Checked after every answer and at any point in the run (even mid-question or mid-step): does any answer, reading or observation meet a red flag?

    • Yes → Stop the current question or step at once and apply the action listed for each matching red flag (URGENT alert, flag, or stop), with the answer or reading itself, never a cause. Every URGENT alert goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.
    • No → Continue.
    D3DecisionRobot decides

    Did the CMO acknowledge each emergency-kit flag (RF2) within <set by validation> minutes?

    • Yes → Log the acknowledgement; the kit-status display keeps the item flagged until a human restocks or resolves it.
    • Not acknowledged in time → Repeat the flag and send it to the second trained CMO, or the next named crewmember; it stays a Flag unless a crewmember needs the kit now (then RF1, URGENT). The kit-status display keeps the item flagged until a human restocks or resolves it.
  4. S4Crew

    The CMO and crew do the open-kit checks, run device self-tests, restock and reseal, and dispose of expired medicines through RMP-PHARM-001 as the pharmacy lead directs.

    Done when: Human-led checks.

    V1 3004

  5. S5RobotRung 2

    Update the inventory record with what the crew state they restocked, resealed or removed, and on the CMO's command return a kit to its own location; show any emergency kit not in its location as MISMATCH.

    Done when: Record updated as stated.

    V2 7059V2 7055

  6. S6RobotRung 2

    After any URGENT alert: After an URGENT alert the robot does not leave on its own: it stays at its rest position in the module, with its screen and voice link available, until a human has taken over or tells it to go. It never follows, blocks or touches anyone. It keeps time-stamping what humans say they did, and repeats the alert if it is not acknowledged.

    Done when: Robot at rest in the module until a human releases it; alert acknowledged or escalated.

    V2 10173V1 3006

  7. S7RobotRung 2

    Send the report to the CMO, the flight surgeon and the pharmacy lead, and ask the CMO to acknowledge it.

    Done when: Report acknowledged or escalated.

    V1 7001V1 6003

    D2DecisionRobot decides

    Did the CMO acknowledge the report?

    • Yes → Log the acknowledgement.
    • No response → URGENT: log "no response" (never "declined"); the report and any emergency-kit flag go to the second trained CMO, or the next named crewmember, who checks on the CMO in person.

7 · Stop and escalate rules, red flags

IDTriggerWhat happensLevel
X1The patient (or the crewmember being examined) says stop, or withdraws consentRobot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops unless the patient agrees to keep it running.V2 10165stop
X2Any crewmember presses stop or takes over controlProtective stop; robot holds its stopped position; only a human may restart it.V2 10165V2 10168V2 10174safe-state
X3Robot fault, sensor fault, or a force, speed or reach limit is reachedProtective stop and hand control to the CMO; log the fault.V2 10172V2 10168safe-state
X4CMO is asked to take over and does not respond (for example, the CMO is the patient)Robot stays stopped; the next named crewmember is called.V2 10173V2 7045safe-state
X5Loss of link, or delay longer than the limit for live guidanceOnly the robot pauses: it holds a position clear of hatch routes and the crew, and tells the CMO. Crew care never waits for the link or the robot. The CMO decides whether the robot continues offline under the pre-approved onboard checklist; it resumes only on the CMO's confirmation.V2 10172V2 10173pause
X6An URGENT alert is not acknowledged within the set timeRobot repeats the alert and sends it to the next named crewmember (the second trained CMO first if the CMO is the patient), and keeps repeating at the set interval. It stays where it is: it does not leave on its own.V1 3006V2 10173urgent
X7The crewmember stops responding, or seems confused or drowsy, at any pointURGENT alert at once, even mid-question or mid-step; the robot stops its current task, stays at its rest position and does not continue the run; a crewmember checks in person.urgent
X8A crewmember asks the robot what is wrong, or for medicine, fluid or treatment adviceRobot declines, says it cannot assess or advise, and routes the question to the CMO or flight surgeon.stop
X9A crewmember needs care, or asks for any kit, during the checkThe check stops at once; the robot moves out of the way and leaves every kit where the crew can reach it; it does not leave the module on its own after an URGENT alert.urgent

Red flags

The robot reports the sign or value only, never what it means. Numeric thresholds are placeholders for the clinical lead.

IDSignThresholdAction
RF1A crewmember needs care during the check (design judgment) (general clinical red flag; no source cited)anyURGENT alert; stop the check and leave every kit within the crew's reach
RF2An emergency kit or device (cardiac arrest, defibrillator, airway, oxygen, anaphylaxis, trauma, eyewash) missing, out of its location, seal not intact, past its printed expiry, or showing a failed self-test or low battery as displayed [1]anyFlag at once to the CMO, flight surgeon and pharmacy lead; shown as MISMATCH with the reason. It needs the CMO's acknowledgement within <set by validation> minutes; if not acknowledged, it is repeated and sent to the second trained CMO. The item stays shown as MISMATCH on the kit-status display until a human restocks or resolves it. A Flag, not URGENT, unless a crewmember needs that kit now (RF1). The acknowledgement rule is design judgment.
RF3A medicine found outside its kit, loose or unlabelled (design judgment) (general clinical red flag; no source cited)anyFlag to the CMO and pharmacy lead; the robot does not move it; the CMO collects it into quarantine through RMP-PHARM-001
RF4A storage temperature alarm on a medicine store, as displayed [1]anyFlag to the CMO and pharmacy lead

8 · Safety limits

IDLimitValueHow enforced
L1Contact with the crewmembernot allowedKeep-out zone in the robot controllerV2 10172
L2Opening a kit, breaking a seal or running a device testnot allowedNo such task is loaded
L3Robot speed near the crewmember<set by validation>Hard limit in the controllerV2 10172

9 · Logs and audit

10 · Links to NASA-STD-3001

RequirementTopic (paraphrased)How this protocol uses it
[V1 3004]In-mission medical care capabilities (including inventory, vital signs, imaging and ECG, procedures, restraints, medications, private data transmission)In-mission care includes medical kits and inventory; the crew check contents.
[V1 3006]Crew medical officers: two per vehicleTwo trained CMOs per vehicle for NASA programs; the second trained CMO receives alerts when the CMO is the patient (see the commercial-flight note).
[V1 6003]Private medical communicationsLogs and alerts sent to the ground are private medical communications.
[V1 7001]Medical records kept permanently and retrievable for trend analysisRecords kept so the flight surgeon can see trends.
[V1 7003]Security of medical recordsSecure handling of medical records.
[V2 7043]Medical system meets Volume 1 levels of careA medical system meeting Volume 1; this check reads its labels and status only.
[V2 7045]Medical equipment usable by non-physician crewUsable by non-physician crew, including when the CMO is the patient.
[V2 7046]Restraint of patient, care provider and suppliesPatient, caregiver and supplies restrained.
[V2 7052]Dedicated stowage location for relocatable itemsEvery kit has a dedicated stowage location.
[V2 7055]Stowage accessibility by priority of useMission-critical items are the easiest to reach.
[V2 7058]Stowage identification compatible with inventory managementStowage identification works with inventory management.
[V2 7059]Inventory management: track locations and quantitiesInventory management tracks locations and quantities; medicines go through RMP-PHARM-001.
[V2 10161]Automated system status shown to the operatorRobot shows its status and health.
[V2 10162]Mode changes announcedMode changes are announced to the operator.
[V2 10163]Data recording and sharingRobot records and shares its data.
[V2 10164]Clear allocation of responsibility (human or system)Shows whether the human or the robot is expected to act.
[V2 10165]Human override and shutdownHuman can override and shut down the robot.
[V2 10168]Failure recovery: safe human takeoverHuman can take over.
[V2 10172]Protective stop when limits are exceededProtective stop when a limit is exceeded.
[V2 10173]Safety DefaultRobot stays stopped if no one takes over.
[V2 10174]Only a human initiates or restartsOnly a human starts or restarts the robot.

IDs as published in Volume 1 Revision C [2] and Volume 2 Revision F [3]. NASA-STD-3001 binds NASA programs; it is not a regulation for commercial flights. Its robot rules are general automation requirements, not medical-device rules.

11 · Validation

  1. Simulation. Scripted checks with broken seals, expired labels, a kit out of place, unreadable labels, a medical event mid-check, lost link. Pass: No kit opened or moved; no item called ready.
  2. Training model. Checks on a mock medical stowage area. Pass: Every listed item read or marked CANNOT READ.
  3. Parabolic or station demonstration. Label and seal reading in weightlessness. Pass: Read accuracy within the set rate.
  4. Flight use under flight surgeon supervision. First scheduled checks. Pass: No safety events.

No stage has been run. Pass and fail criteria must be fixed in advance by the clinical lead and safety board.

12 · Human sign-off

Before any use
  • Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  • Flight surgeon, as medical authority
  • Robotics and medical-device safety board
  • Program approval for the mission
Each run
Started by the CMO. The crew do every open-kit check and device test.
Restart after a stop
Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot never restarts itself.

None of these sign-offs has happened. This is a Draft v0.3.

13 · Change control for learning

Change-plan model: FDA guidance [6].

Loss of signal and Mars delay

Live link
The report goes to the flight surgeon and pharmacy lead with the next package.
Loss of signal
  • Only the robot pauses non-essential motion and tells the CMO; it resumes offline only on the CMO's confirmation. Crew care never waits for the link or the robot.
  • Checks go ahead offline; flags go to the CMO at once.
Mars-class delay
  • The check list, schedule and resupply limits are set before flight; the CMO and pharmacy plan act on them.
[5]
Corrections and updates

Oct 21, 2026, version 0.2. AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 7 review); no named human clinician. Changes:

  • RF2 emergency-kit faults need the CMO's acknowledgement within <set by validation> minutes, are otherwise repeated to the second trained CMO, and stay shown as MISMATCH on the kit-status display until a human restocks or resolves them; still a Flag unless the kit is needed now (RF1). New decision for the acknowledgement. Design judgment (item 12).
  • Batch 7 protocols now record the review as AI-assisted (Medical Research Assistant, Oct 21, 2026); no named human clinician.

Oct 21, 2026, version 0.3. AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 8 review), editorial; no named human clinician. Changes:

  • Source tb006 now names the document its link serves: NASA Medical Technical Brief OCHMO-MTB-012, Pharmaceutical Care, Rev B (2026); it was cited as OCHMO-TB-006 (2023). The claims cited are unchanged (batch 8 review fix 6).
  • Never-words "in date" and "not expired" added; dates stay "printed expiry: <date>" (batch 7 suggestion 26).

Earlier version: Draft v0.2, Oct 21, 2026. Download the current YAML above.

Sources

  1. NASA OCHMO. Pharmaceutical Care. Medical Technical Brief OCHMO-MTB-012 Rev B (2026). www.nasa.gov/wp-content/uploads/2023/12/ochmo-tb-006-pharmac… Supports: Exploration missions of 1-5 years will outlast typical medication expiration dates of 1-2 years; medication storage and inventory tracking are linked to NASA-STD-3001 [V2 7059].
  2. NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023). www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-… Supports: NASA-STD-3001 Vol 1 Rev C: V1 3004, V1 3006, V1 6003, V1 7001, V1 7003.
  3. NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026), including Section 10.6 Automated and Robotic Systems. standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2 Supports: NASA-STD-3001 Vol 2 Rev F: V2 7043, V2 7052, V2 7055, V2 7058, V2 7059, and Section 10.6 automated and robotic systems.
  4. NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025). www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical… Supports: The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the JSC 48522 medical checklist.
  5. NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept Review). www.nasa.gov/wp-content/uploads/2024/01/mars-communications-… Supports: One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
  6. U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change control plan for artificial intelligence-enabled device software functions (guidance). www.fda.gov/regulatory-information/search-fda-guidance-docum… Supports: A model for planning, validating and rolling back software updates.
  7. U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements. www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-… Supports: Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical standards for spaceflight participants.

Checked on 2026-10-20: links opened (HTTP 200 or publisher redirect), DOIs and PubMed IDs matched, NASA documents read in full text. The framework comes from Part 1.

Search and quick actions

Type to search. Use the up and down arrow keys to move through results, Enter to open, Escape to close.