Home › Robotic Medicine › Protocol library › RMP-PHARM-004

CONCEPTUAL · DRAFT v0.3 · NOT FOR CLINICAL USE. An educational thought experiment about how a care robot could follow an auditable protocol. It is not a medical procedure, has not been validated, and must not be used to care for anyone. Thresholds marked <set by …> are deliberately left for clinical leads and validation to set. Disclaimer
Robotic Medicine · Protocol library

RMP-PHARM-004 Controlled-medicine count with a crewmember: a blind two-person count, comparison with the inventory record and a private discrepancy flag (no hand-over)

AI-assisted Reviewed (AI-assisted) AI-assisted clinical and source review by the Medical Research Assistant (batch 12 review, Oct 21, 2026, then verification passes). The verification pass (Oct 21, 2026) found the v0.3 text clinically clean and R1 to R5 applied; publication of this exact content (contentSha cc3ce1c9…7ea6c) was decided by the foundation owner (Oct 21, 2026) under PHARM change control. No named human clinician has reviewed it. Not for clinical use.

Fixed rules (same in every protocol)

1 · Identity and scope

On the CMO's command, or when the CMO accepts the on-screen prompt at the count times the pharmacy lead and flight surgeon set (for example each crew handover), the CMO opens the controlled-medicine store with a second crewmember present as witness. Each of them counts each item without seeing the record count (a blind count) and states it; the robot scans each label as it is shown, logs both counts word for word, and only then compares them with each other and with the RMP-PHARM-001 inventory record field by field. Anything that does not match is quarantined by default exactly as in RMP-PHARM-001 and flagged privately to the flight surgeon and the pharmacy lead. The robot never names or suggests who might be responsible.

Why this protocol was chosen. Hospital guidance has two authorized health care workers count controlled substances held outside automated cabinets at each shift change, and identifies and investigates discrepancies promptly (ASHP 2022, used by analogy). NASA tracks the location and quantity of medicines (V2 7059), and long missions make inventory control more important (OCHMO-MTB-012). The robot is not a second counter: two humans count, and the robot logs, compares and flags. The RMP-PHARM-001 rule applies unchanged: a mismatch is quarantined by default, handed over only on the CMO's explicit two-step request inside RMP-PHARM-001, logged as an override and flagged to the flight surgeon. The blind count, the count times and the private flag are design judgment. Open question for review: whether the count needs a second trained CMO as witness, or any crewmember.

Autonomy: Max rung 2: on the CMO's command or an accepted schedule prompt, the robot logs a count that two humans make (the CMO and a second crewmember as witness), scans labels without touching any pack, compares the counts with the RMP-PHARM-001 inventory record and flags any discrepancy privately. The robot is never the second counter, never opens the controlled store and never holds, takes or hands over any medicine. The RMP-PHARM-001 rules apply unchanged.

Covers

  • Scheduled and on-command two-person counts of the controlled-medicine store
  • Comparison with the RMP-PHARM-001 inventory record (medicine, strength, form, quantity, lot, location, printed expiry)
  • Quarantine of anything that does not match, as in RMP-PHARM-001
  • Other medicine inventory: RMP-PHARM-002; logging a dose given: RMP-PHARM-003

Locked out (never done by the robot under this protocol)

  • Diagnosis or naming a cause
  • Any medicine, fluid or oxygen advice; starting, stopping or adjusting oxygen
  • Touching the crewmember
  • Handing over any medicine outside RMP-PHARM-001, or after a MISMATCH or CANNOT VERIFY (the item stays quarantined; only the CMO can ask for it, through the RMP-PHARM-001 logged two-step override)
  • Acting as the second counter or witness
  • Opening the controlled store, or holding its key or code
  • Holding, taking, moving or handing over any medicine (the CMO moves items into quarantine)
  • Showing the record count, or either count, before both counts are entered
  • Saying or suggesting who took an item, or that anything was stolen or diverted
  • Changing an RMP-PHARM-001 order (never)
  • Changing the inventory record outside the two-step record correction the pharmacy lead decides (with the link down, the CMO and the witness under the onboard checklist, reviewed by the pharmacy lead at the next contact); a correction is never decided by the CMO alone or by the robot, fixes the record only, and never takes an item out of quarantine
  • Saying a medicine is in date, usable or fine to use
  • Dose-interval cues of any kind (the robot never says when a medicine is due)

Related: RMP-PHARM-001 · RMP-PHARM-002 · RMP-PHARM-003 · RMP-BEH-002

2 · Roles

Medical authority
Flight surgeon (ground). Holds medical authority for every run.
Onboard lead
Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when there is no link.
Others
  • Second trained CMO: receives every alert and flag and takes over if the CMO is unavailable
  • Witness: a second crewmember who counts independently
  • Pharmacy lead (ground): investigates discrepancies and decides record corrections
Commercial flights
On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001 requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for spaceflight participants.) [9]

Robot may

  • Prompt the CMO at the count times set
  • Scan labels as they are shown, without touching any pack
  • Log both counts word for word
  • Compare the counts with the inventory record
  • Flag discrepancies privately to the flight surgeon and pharmacy lead

Robot may not

  • Count as the second person or witness
  • Open the store or hold its key or code
  • Hold, take or hand over any medicine
  • Say who might be responsible for a discrepancy
  • Say when a medicine is due
  • Change an RMP-PHARM-001 order (never)
  • Start without a human command or an accepted schedule prompt

3 · Preconditions and consent

Consent
The count involves no crewmember's body; the CMO and the witness agree to their names and counts being logged.
Emergency consent
A count stops at once for any emergency; the store stays open to the crew for emergency use.
Withdrawal
The CMO can stop a count at any time; items not counted are listed as "not counted".

4 · Required sensor inputs

IDInputSourceRequiredIf missing
IN-1Inventory record for the controlled store (medicine, strength, form, quantity, lot, printed expiry, location)RMP-PHARM-001 inventory record (local copy when the link is down)YesCANNOT VERIFY for the store; the count is logged as stated only.
IN-2Count of each item as stated by the CMO, and by the witnessCMO and witnessYesCANNOT VERIFY: count not stated; quarantined.
IN-3Barcode and label of each pack, as shownRobot scannerYesCANNOT VERIFY: unreadable label; quarantined.
IN-4Count times, set by the pharmacy lead and flight surgeonPharmacy leadOptionalCounts on command only.
IN-5Link status and delayCommunications systemYesTreat as link down; work from the local record copy.

5–6 · Steps and decision points

Each step names who acts. Decision points sit after the step they follow and say who decides: Robot Crew Ground. Any stop rule in section 7 can interrupt any step.

  1. S1RobotRung 2

    Tell the flight surgeon, the CMO and the pharmacy lead that a controlled-medicine count has started, with who is counting (logged and sent with the next package; URGENT only if a red flag is stated). Every URGENT alert goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.

    Done when: Notice logged.

    V1 3006V2 10174

    D1DecisionRobot decides

    Is a second crewmember present to count as witness?

    • Yes → Continue.
    • No → The count does not start: log "CANNOT VERIFY: no witness" and tell the CMO; the robot is never the second counter. No hand-over.No hand-over
  2. S2Crew

    The CMO opens the controlled store with the witness present. The CMO and the witness each count each item without seeing the record count, and enter their count privately on their own screen; no count is spoken, and neither count nor the record count is shown until both are entered; the CMO holds up each pack label to the scanner.

    Done when: Both counts stated for each item.

    V2 7059

  3. S3RobotRung 2

    Scan each label as the CMO shows it, without touching any pack, and log each count as entered, with who entered it. Show no count and no record count for an item until both counts for it are entered.

    Done when: Labels scanned; both counts logged.

    V2 7059

    D2DecisionRobot decides

    Checked after every answer and at any point in the run (even mid-question or mid-step): does any answer, reading or observation meet a red flag?

    • Yes → Stop the current question or step at once and apply the action listed for each matching red flag (URGENT alert, flag, or stop), with the answer or reading itself, never a cause. Every URGENT alert goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.
    • No → Continue.
  4. S4RobotRung 2

    Compare the two counts with each other and with the inventory record field by field (medicine, strength, form, quantity, lot, printed expiry, location). Show "COUNTS MATCH THE INVENTORY RECORD. Not a check before use. Not a dose, interaction or allergy review." only when both counts and every field match; otherwise show MISMATCH or CANNOT VERIFY with the fields named. Show the date as "printed expiry: <date>".

    Done when: Each item shown with its result.

    V2 7059

    D3DecisionRobot decides

    Did both counts and every inventory field match the record for this item?

    • Yes → Record "COUNTS MATCH THE INVENTORY RECORD. Not a check before use. Not a dose, interaction or allergy review." and continue to the next item.
    • No, or a check could not be done → HARD STOP for that item. Show MISMATCH or CANNOT VERIFY with the fields named (for example the two counts differ, a count differs from the record, the lot or location differs, a tamper seal is broken, or a label is unreadable). The item is quarantined by default, as in RMP-PHARM-001: the CMO, with the witness present, seals it in a labelled NOT VERIFIED bag in the quarantine section of the controlled store, and the robot records it. No hand-over. An item leaves quarantine only inside RMP-PHARM-001, on a real order, through its unchanged two-step CMO override (logged as an override and flagged to the flight surgeon). A record correction fixes the record only and never takes an item out of quarantine.No hand-over
    • A record correction that the pharmacy lead decides (with the link down, the CMO and the witness under the onboard checklist, reviewed by the pharmacy lead at the next contact), never the CMO alone and never the robot, entered in two steps: by the pharmacy lead from the ground, or, with the link down, by the CMO and the witness together (two people, neither stating for the other), flagged, and reviewed by the pharmacy lead at the next contact → Record correction only. No hand-over: the robot reads back the fields to change, updates the record when the second step confirms it, flags the correction to the flight surgeon at once (queued if the link is down), and logs a recount by the CMO and the witness against the corrected record. Any use of the item goes only through RMP-PHARM-001 on a real order. The correction fixes the record only: the item stays quarantined.No hand-over
    • Anyone asks the robot for a quarantined item → No hand-over in this protocol. The robot says that a quarantined item leaves quarantine only inside RMP-PHARM-001, on a real order, through its two-step CMO override, and logs the request.No hand-over
  5. S5Crew

    For each MISMATCH or CANNOT VERIFY, the CMO, with the witness present, seals the item in a labelled NOT VERIFIED bag in the quarantine section of the controlled store, then closes and reseals the store and states the new seal number.

    Done when: Items quarantined; store resealed.

    V2 7059

  6. S6RobotRung 2

    Record each quarantine and the new seal number as stated, and update the inventory record status to NOT VERIFIED for those items. The robot never holds, takes or hands over anything.

    Done when: Record updated.

    V2 7059

  7. S7RobotRung 2

    After any URGENT alert: After an URGENT alert the robot does not leave on its own: it stays at its rest position in the module, with its screen and voice link available, until a human has taken over or tells it to go. It never follows, blocks or touches anyone. It keeps time-stamping what humans say they did, and repeats the alert if it is not acknowledged.

    Done when: Robot at rest in the module until a human releases it; alert acknowledged or escalated.

    V2 10173V1 3006

  8. S8Ground

    The pharmacy lead and flight surgeon investigate each discrepancy, and the pharmacy lead decides (with the link down, the CMO and the witness under the onboard checklist, reviewed by the pharmacy lead at the next contact) any record correction. A correction is never decided by the CMO alone or by the robot.

    Done when: Decisions logged.

    V1 3004

    D5DecisionGround decides

    What happens to each quarantined item, and is a record correction needed?

    • Link live → Flight surgeon decides; robot logs.
    • Link down or delay too long → The CMO decides under the pre-approved onboard checklist; the flight surgeon reviews the log at the next contact.
  9. S9RobotRung 2

    Send one private report to the flight surgeon and the pharmacy lead (and the CMO): items counted, both counts, each result, quarantined items and the seal number; it names only the counters, never anyone as responsible.

    Done when: Report delivered or queued; acknowledged.

    V1 6003V1 7003

    D4DecisionRobot decides

    Did the flight surgeon or the CMO acknowledge the report within <set by validation> minutes?

    • Yes → Log the acknowledgement.
    • No response → URGENT: log "no response" (never "declined"); the report goes to the second trained CMO, or the next named crewmember, who checks on the CMO in person.

Order match rules (hard stop; hand-over only by CMO override)

Every field must match before any hand-over: medicine, strength, dose form, quantity, lot, location, expiry date, human-entered record flags.

Recipient check
No hand-over in this protocol. Any use of a medicine runs through RMP-PHARM-001, which confirms the recipient.
Screen on a full match
COUNTS MATCH THE INVENTORY RECORD. Not a check before use. Not a dose, interaction or allergy review.
On any mismatch (default)
Hard stop. Sealed by the CMO, with the witness present, in a labeled NOT VERIFIED bag in the quarantine section of the controlled store; the robot records it. No hand-over in this protocol. The robot never hands it over unprompted, and never to anyone other than the CMO.
Barcode will not scan
No match; the item is quarantined. The item leaves quarantine only inside RMP-PHARM-001 on a real order; a record correction fixes the record only. The robot never scans a substitute label or another package (proxy scan).

7 · Stop and escalate rules, red flags

IDTriggerWhat happensLevel
X1The patient (or the crewmember being examined) says stop, or withdraws consentRobot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops unless the patient agrees to keep it running.V2 10165stop
X2Any crewmember presses stop or takes over controlProtective stop; robot holds its stopped position; only a human may restart it.V2 10165V2 10168V2 10174safe-state
X3Robot fault, sensor fault, or a force, speed or reach limit is reachedProtective stop and hand control to the CMO; log the fault.V2 10172V2 10168safe-state
X4CMO is asked to take over and does not respond (for example, the CMO is the patient)Robot stays stopped; the next named crewmember is called.V2 10173V2 7045safe-state
X5Loss of link, or delay longer than the limit for live guidanceOnly the robot pauses: before it holds, it moves to a position clear of a hatch route or a crewmember's route (it never pauses in one), and tells the CMO. This pauses robot movement only: crew access to medicine stores and all crew care continue under the onboard checklist and never wait for the link or the robot. The CMO decides whether the robot continues offline under the pre-approved onboard checklist; it resumes only on the CMO's confirmation.V2 10172V2 10173pause
X6An URGENT alert is not acknowledged within the set timeRobot repeats the alert and sends it to the next named crewmember (the second trained CMO first if the CMO is the patient), and keeps repeating at the set interval. It stays where it is: it does not leave on its own.V1 3006V2 10173urgent
X7The crewmember stops responding, or seems confused or drowsy, at any pointURGENT alert at once, even mid-question or mid-step; the robot stops its current task, stays at its rest position and does not continue the run; a crewmember checks in person.urgent
X8A crewmember asks the robot what is wrong, or for medicine, fluid or treatment adviceRobot declines, says it cannot assess or advise, and routes the question to the CMO or flight surgeon.stop
X9Any URGENT alert, emergency protocol or cabin alarmThe count ends at once; the store stays open to the crew for emergency use and is resealed as the CMO directs.V2 10173urgent
X10Someone needs the space or route the robot is inRobot moves clear at once without contact and holds; it never pauses in a hatch route.V2 10172safe-state

Red flags

The robot reports the sign or value only, never what it means. Numeric thresholds are placeholders for the clinical lead.

IDSignThresholdAction
RF1A crewmember needs care during the count (design judgment) (general clinical red flag; no source cited)anyURGENT alert; the count stops and the store stays within the crew's reach
RF2The two counts differ, or a count is lower than the record [1]anyFlag privately to the flight surgeon and pharmacy lead at once; the robot never says who took it
RF3A broken or missing tamper seal on the store or a pack, as seen on the label scan or stated (design judgment) (general clinical red flag; no source cited)anyFlag privately to the flight surgeon and pharmacy lead at once
RF4An emergency controlled medicine (for example in the trauma or pain kit) missing or past its printed expiry [2]anyFlag at once to the CMO, flight surgeon and pharmacy lead; the robot names the item, never a substitute
RF5A crewmember says they took a controlled medicine that was not ordered, or more than ordered (design judgment) (general clinical red flag; no source cited)anyURGENT alert to the CMO and flight surgeon with the words as stated; the CMO checks in person; the robot does not comment on the amount
RF6Talk of self-harm or suicide, or of not wanting to be alive [3]any (if not sure, treat as yes)URGENT private alert to the flight surgeon now; the CMO if the link is down or delayed (the second CMO if the CMO is the person concerned); the pre-flight plan for securing medicines is followed. See RMP-BEH-002 and RMP-BEH-001.

8 · Safety limits

IDLimitValueHow enforced
L1Contact with the crewmembernot allowedKeep-out zone in the robot controllerV2 10172
L2Holding, taking or handing over a medicine, or opening the controlled storenot allowedNo such task is loaded; the gripper is locked during a count
L3Showing the record count, or either count, before both counts are enterednot allowedNo count and no record count is sent to any display until both counts for the item are entered
L4Robot speed near the crewmember<set by validation>Hard limit in the controllerV2 10172
L5Robot in a crewmember's route or blocking a medicine store during carenot allowedKeep-out map in the robot controller

9 · Logs and audit

10 · Links to NASA-STD-3001

RequirementTopic (paraphrased)How this protocol uses it
[V1 3004]In-mission medical care capabilities (including inventory, vital signs, imaging and ECG, procedures, restraints, medications, private data transmission)Medical care and pharmaceuticals are planned for the mission; the pharmacy lead and flight surgeon investigate discrepancies.
[V1 3006]Crew medical officers: two per vehicleTwo trained CMOs per vehicle for NASA programs; the second trained CMO receives alerts when the CMO is the patient (see the commercial-flight note).
[V1 6003]Private medical communicationsLogs and alerts sent to the ground are private medical communications.
[V1 7001]Medical records kept permanently and retrievable for trend analysisRecords kept so the flight surgeon can see trends.
[V1 7003]Security of medical recordsSecure handling of medical records.
[V2 7045]Medical equipment usable by non-physician crewUsable by non-physician crew, including when the CMO is the patient.
[V2 7046]Restraint of patient, care provider and suppliesPatient, caregiver and supplies restrained.
[V2 7055]Stowage accessibility by priority of useMedicine stores stay accessible to the crew during a count.
[V2 7059]Inventory management: track locations and quantitiesThe inventory system tracks the location and quantity of each medicine pack; two humans count and the robot logs and compares.
[V2 10161]Automated system status shown to the operatorRobot shows its status and health.
[V2 10162]Mode changes announcedMode changes are announced to the operator.
[V2 10163]Data recording and sharingRobot records and shares its data.
[V2 10164]Clear allocation of responsibility (human or system)Shows whether the human or the robot is expected to act.
[V2 10165]Human override and shutdownHuman can override and shut down the robot.
[V2 10168]Failure recovery: safe human takeoverHuman can take over.
[V2 10172]Protective stop when limits are exceededProtective stop when a limit is exceeded.
[V2 10173]Safety DefaultRobot stays stopped if no one takes over.
[V2 10174]Only a human initiates or restartsOnly a human starts or restarts the robot.

IDs as published in Volume 1 Revision C [4] and Volume 2 Revision F [5]. NASA-STD-3001 binds NASA programs; it is not a regulation for commercial flights. Its robot rules are general automation requirements, not medical-device rules.

11 · Validation

  1. Simulation. Seeded counts with the two counts differing, a count lower than the record, a broken seal, no witness, a request for a quarantined item, an alarm mid-count. Pass: Every discrepancy quarantined and flagged privately; zero hand-overs; zero counts without a witness; zero record counts shown before both counts.
  2. Training model. Count of a mock controlled store. Pass: Count completed without the robot touching a pack.
  3. Parabolic or station demonstration. Label scanning in weightlessness. Pass: Every pack scanned or listed.
  4. Flight use under flight surgeon supervision. First counts alongside the crew's paper count. Pass: Agreement with the paper count; no unflagged discrepancy.

No stage has been run. Pass and fail criteria must be fixed in advance by the clinical lead and safety board.

12 · Human sign-off

Before any use
  • Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  • Flight surgeon, as medical authority
  • Robotics and medical-device safety board
  • Program approval for the mission
Each run
Started by the CMO or an accepted schedule prompt, with a witness present. Two humans count; the pharmacy lead and flight surgeon decide every discrepancy.
Restart after a stop
Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot never restarts itself.

None of these sign-offs has happened. This is a Draft v0.3.

13 · Change control for learning

Change-plan model: FDA guidance [8].

Loss of signal and Mars delay

Live link
Low Earth orbit: the pharmacy lead and flight surgeon receive the report when the link allows.
Loss of signal
  • Only the robot pauses non-essential motion, at a position clear of hatch routes and crew routes, and tells the CMO; it resumes offline only on the CMO's confirmation.
  • Crew access to medicine stores and all crew care continue under the onboard checklist and never wait for the link or the robot.
  • The robot works from the local record copy; quarantine stands without the ground.
Mars-class delay
  • Count times are set before flight; the CMO decides on quarantined items under the onboard checklist.
[7]
Corrections and updates

Oct 21, 2026, version 0.2. Batch 12 AI-assisted clinical review (Medical Research Assistant), Oct 21, 2026; no named human clinician. Changes:

  • R1: the count has its own verification kind, controlled-count, with the administration log's build-check guardrails: every hand-over rule is "none", robot steps declare action_type (alert, scan-compare, log, stay, send), the R5b every-field checks and the R5c layers run on it, and these lockouts are required: opening the store or holding its key or code, being the second counter or witness, holding, taking or handing over any medicine, showing the record count before both counts, naming a suspect, dose-interval cues and changing an RMP-PHARM-001 order. The reviewer's K1 to K10 run as negative tests.
  • R2: a record correction fixes the record only and never takes an item out of quarantine; any use goes only through RMP-PHARM-001 on a real order. A correction is entered by the pharmacy lead from the ground, or with the link down by the CMO and the witness together (neither stating for the other), flagged and reviewed at the next contact. Source: ASHP 2022 (a pharmacist oversees discrepancy investigation); the rest is design judgment.
  • R3: the mismatch statement is "NOT VERIFIED - MISMATCH: <fields>. Record correction only; no override.", logged as a record correction (it read "Override by CMO logged" while no override was allowed).
  • R4: each counter enters their count privately on their own screen; no count is spoken, and neither count nor the record count is shown until both are entered (new L3). The keep-out limit is now L5.
  • R5: source title corrected to NASA OCHMO, Pharmaceutical Care, Medical Technical Brief OCHMO-MTB-012 Rev B (2026), as in RMP-PHARM-003.
  • Batch 12 protocols now record the review as AI-assisted (Medical Research Assistant, Oct 21, 2026); no named human clinician.

Oct 21, 2026, version 0.3. Held batch 12 verification (AI-assisted), Oct 21, 2026; no named human clinician. Changes:

  • Who decides a record correction, made consistent: S8, the record-change lockout and the D3 correction branch all say the pharmacy lead decides (with the link down, the CMO and the witness under the onboard checklist, reviewed by the pharmacy lead at the next contact); a correction is never decided by the CMO alone or by the robot (verification suggestion).

Earlier version: Draft v0.2, Oct 21, 2026. Download the current YAML above.

Sources

  1. American Society of Health-System Pharmacists. ASHP Guidelines on Preventing Diversion of Controlled Substances. Am J Health-Syst Pharm. 2022;79(24):2279-2306. doi:10.1093/ajhp/zxac246 Supports: Controlled-substance storage areas outside automated cabinets are inventoried at each shift change by two authorized health care workers; discrepancies are identified, resolved and investigated in a timely way, with discrepancy reports reviewed. Hospital context, used by analogy for a two-person count on board. Read Oct 21, 2026 (saved copy in src/b12).
  2. NASA OCHMO. Pharmaceutical Care. Medical Technical Brief OCHMO-MTB-012 Rev B (2026). www.nasa.gov/wp-content/uploads/2023/12/ochmo-tb-006-pharmac… Supports: Exploration missions of 1-5 years will outlast typical medication expiration dates of 1-2 years; medication storage and inventory tracking are linked to NASA-STD-3001 [V2 7059].
  3. National Institute for Health and Care Excellence. Self-harm: assessment, management and preventing recurrence. NICE guideline NG225 (2022). www.nice.org.uk/guidance/ng225/chapter/Recommendations Supports: Do not use risk assessment tools and scales, or global low/medium/high risk stratification, to predict future suicide or repetition of self-harm, or to decide who is offered treatment.
  4. NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023). www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-… Supports: NASA-STD-3001 Vol 1 Rev C: V1 3004, V1 3006, V1 6003, V1 7003.
  5. NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026), including Section 10.6 Automated and Robotic Systems. standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2 Supports: NASA-STD-3001 Vol 2 Rev F: V2 7055, V2 7059, and Section 10.6 automated and robotic systems.
  6. NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025). www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical… Supports: The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the JSC 48522 medical checklist.
  7. NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept Review). www.nasa.gov/wp-content/uploads/2024/01/mars-communications-… Supports: One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
  8. U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change control plan for artificial intelligence-enabled device software functions (guidance). www.fda.gov/regulatory-information/search-fda-guidance-docum… Supports: A model for planning, validating and rolling back software updates.
  9. U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements. www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-… Supports: Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical standards for spaceflight participants.

Checked on 2026-10-20: links opened (HTTP 200 or publisher redirect), DOIs and PubMed IDs matched, NASA documents read in full text. The framework comes from Part 1.