# RMP-RENAL-002 Trouble passing urine: bladder ultrasound support with the crew holding the probe, relay to the flight surgeon (diagnostic only)
# Version 0.6 (Draft). CONCEPTUAL. NOT FOR CLINICAL USE.
# Space Medicine Foundation robot-readable protocol library. Format: https://spacemedicinefoundation.org/robotic-medicine/protocols/schema/
# Every threshold marked <set by ...> must be set by the named owner (see the schema) before any use.
schema_version: '0.3'
version: '0.6'
status: draft
conceptual: true
not_for_clinical_use: true
max_autonomy: 2
missions:
- leo
- commercial
- moon
- mars
updated: '2026-10-21'
review:
  state: reviewed-ai-assisted
  requested: '2026-10-20'
  reviewed: '2026-10-20'
  human_clinician_review: false
  text: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 20, 2026 (batch 2 review,
    51 corrections); all corrections are applied in this version. No named human clinician has reviewed
    it.
fixed_rules:
  flight_surgeon_is_medical_authority: true
  cmo_acts_under_delegated_care: true
  human_starts_and_restarts_robot: true
  robot_never_diagnoses: true
  no_drug_or_fluid_advice_by_robot: true
  safe_state_on_fault_or_no_response: true
  lost_link_pause_then_cmo_decides: true
robot_lockouts:
  diagnosis: true
  drug_or_fluid_advice: true
  starting_stopping_or_adjusting_oxygen_or_any_medicine: true
  self_start_or_self_restart: true
  changing_own_protocol_thresholds_or_limits: true
  handing_over_a_mismatched_item_as_verified: true
id: RMP-RENAL-002
slug: rmp-renal-002
title: >-
  Trouble passing urine: bladder ultrasound support with the crew holding the probe, relay to the flight
  surgeon (diagnostic only)
category: renal
tier: 1
rung_note: >-
  In this draft the CMO holds the ultrasound probe. The robot fetches, holds the screen, shows the scan
  guide and records. Robot-held probe scanning, as in RMP-RENAL-001, is left out of this protocol until
  it has been validated for this task.
result_language:
  use:
  - device estimate, for human review
  - images acquired, for human review
  - non-diagnostic (quality too low)
  never:
  - normal
  - no problem found
  - all clear
  - negative
  - any diagnosis or cause
  - bladder empty
  - bladder full
  - no retention
  - catheter needed
  - catheter not needed
  - safe
scope:
  summary: >-
    When a crewmember cannot pass urine, passes only small amounts, or feels the bladder has not emptied,
    the robot alerts the CMO and flight surgeon, asks red-flag questions first, brings the ultrasound
    device and the sealed catheter kit (unopened), turns its camera away, holds the screen where the CMO
    can see it, shows the pre-loaded bladder scan guide, and records the images and the device's bladder-volume
    estimate for the flight surgeon. The CMO holds the probe. Whether to catheterize, and any medicine,
    are human decisions. The robot never says the bladder is full or empty.
  why: >-
    Urinary retention can happen in the first days of flight. NASA lists fluid shifts, a changed feeling
    of bladder fullness and changes in the voiding reflex among the contributors, and treats it with fluid
    management and intermittent catheterization as needed. US astronaut records list 9 events. In one
    reported case, an astronaut who had taken promethazine and scopolamine before launch could not pass
    urine on reaching orbit, had intermittent catheterization for the first 3 days and then a catheter
    left in place (supplies were short) for a total of 4 days, could pass urine again after day 7, had
    retention return on day 10 (1 day before landing), and had it recur for the first 24 hours of the
    next mission. Retention with back pain or saddle numbness can also signal cauda equina syndrome; a
    bladder scan must not be used on its own to decide, and a low volume does not rule out nerve compression.
  covers:
  - >-
    Unable to pass urine, passing only small amounts, or feeling the bladder has not emptied, in a crewmember
    who is awake
  - Repeat scans when the flight surgeon asks, including after a catheter is removed
  locked_out:
  - >-
    Inserting, handling or removing a urinary catheter (the CMO does it; RMP-RENAL-005 covers supplies
    and hygiene only)
  - Opening the sealed catheter kit
  - Holding or moving the ultrasound probe (in this version)
  - Deciding whether to catheterize, or what volume is too much
  - Any drug or fluid advice, including whether to stop a medicine
  - Saying the bladder is full or empty, or naming a cause
  - >-
    Photographing or filming the patient; the robot camera faces away and stops recording while the patient
    is exposed
  - Therapeutic ultrasound modes
  related:
  - RMP-RENAL-001
  - RMP-RENAL-004
  - RMP-RENAL-005
  - RMP-MSK-004
  - RMP-NEURO-004
  - RMP-INF-001
  - RMP-PHARM-001
roles:
  medical_authority: Flight surgeon (ground). Holds medical authority for every run.
  onboard_lead: >-
    Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when
    there is no link.
  others:
  - 'Patient: can stop at any time and chooses who is present'
  - 'Ground ultrasound expert: can guide the CMO live when the link allows'
  robot_may:
  - >-
    Fetch the ultrasound device, gel, the sealed catheter kit (unopened) and a privacy screen, and restrain
    them by the patient
  - Turn its camera away and stop recording while the patient is exposed
  - Hold the ultrasound screen where the CMO can see it and show the pre-loaded scan guide
  - >-
    Check the catheter kit's seal and expiry, and that antiseptic and spare catheters are in stock, before
    bringing it
  - >-
    Record the images and the device's own volume estimate (before and after passing urine, if the crewmember
    can), labeled "device estimate, for human review"
  - >-
    Show the CMO and flight surgeon the human-entered dose log so the flight surgeon can review medicines
    that can contribute to retention (shown only; the robot names no medicine and gives no advice)
  - Send everything to the flight surgeon; bag waste and clean equipment
  robot_may_not:
  - Hold or move the probe, or touch the patient
  - Open the catheter kit or handle a catheter
  - Interpret images or say what the volume means
  - Suggest catheterization, medicines or fluids
  - Film or photograph the patient
  - Start without a human command
  commercial: >-
    On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority
    and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001
    requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for
    spaceflight participants.)
preconditions:
- id: PRE-1
  text: >-
    A crewmember started the run by command; the CMO is present and trained in bladder ultrasound and
    catheterization.
  check: human_start == true and cmo_present == true
  refs:
  - V2 10174
  - V2 10168
- id: PRE-2
  text: Patient consent recorded; the patient chooses who is present.
- id: PRE-3
  text: Patient, CMO, device and kits restrained.
  refs:
  - V2 7046
- id: PRE-4
  text: Ultrasound device self-test passed and gel available; robot self-test passed.
  refs:
  - V2 10161
- id: PRE-5
  text: Robot camera turned away and recording off before the patient is exposed.
  check: robot_camera_recording == false
- id: PRE-6
  text: Link status and current delay known.
consent:
  standard: The patient agrees by voice or touch to the questions, the scan and the robot being present;
    logged.
  emergency: >-
    If the patient cannot agree, this protocol is not used; the CMO and flight surgeon decide care directly.
  withdrawal: The patient can stop at any time; the robot backs away.
inputs:
- id: IN-1
  name: >-
    Red-flag answers: fever or chills; back pain with leg weakness or numbness, or numbness between the
    legs; blood in urine; flank pain; lower belly pain or urge; severe pain; a recent catheter with fever,
    burning or cloudy urine
  source: Patient, by voice or touch
  required: true
  if_missing: Ask the CMO to ask in person; never treated as "no".
- id: IN-2
  name: Time of the last urine passed and how much (as reported)
  source: Patient
  required: true
  if_missing: Log "not answered".
- id: IN-3
  name: Bladder ultrasound images with the device quality score
  source: Ultrasound device, probe held by the CMO
  required: true
  if_missing: Label "non-diagnostic (quality too low)"; never report a volume.
- id: IN-4
  name: >-
    Bladder-volume estimate calculated by the device (if it has that function), before and, if the crewmember
    can pass some urine, after
  source: Ultrasound device
  required: false
  if_missing: Log "no device estimate"; images still go to the flight surgeon.
- id: IN-5
  name: >-
    Medicines given in the past <set by flight surgeon> hours, as logged by humans (shown so the flight
    surgeon can review medicines that can contribute to retention; the robot names no medicine and does
    not advise)
  source: Medical record or dose log
  required: false
  if_missing: Note "no dose log available".
- id: IN-6
  name: Catheter kit seal and expiry, antiseptic and spare catheters in stock
  source: Inventory system
  required: true
  if_missing: Report the shortfall to the CMO before the kit is needed.
- id: IN-7
  name: Time since launch
  source: Mission clock
  required: true
  if_missing: Ask the CMO.
- id: IN-8
  name: Link status and delay
  source: Communications system
  required: true
  if_missing: Treat as link down.
steps:
- id: S1
  actor: robot
  rung: 2
  action: >-
    On a crew command, alert the CMO and, if the link is up, send a "trouble passing urine" alert to the
    flight surgeon (queued if the link is down).
  success: CMO acknowledges; alert delivered or queued.
  refs:
  - V2 10174
  - V1 6003
- id: S2
  actor: robot
  rung: 2
  action: >-
    Ask the red-flag questions first, then the time and amount of the last urine passed; log the answers
    word for word.
  success: Answers logged, or marked for the CMO to ask.
- id: S3
  actor: robot
  rung: 2
  action: >-
    Bring the ultrasound device, gel, the sealed catheter kit (unopened, after checking its seal and expiry
    and that antiseptic and spare catheters are in stock) and a privacy screen; restrain them by the patient.
    Report any shortfall to the CMO.
  success: Items restrained at the patient; any shortfall reported.
  limit: <set by validation>
  refs:
  - V2 7046
- id: S4
  actor: robot
  rung: 2
  action: Turn the robot camera away from the patient and stop recording; show the camera-off indicator.
  success: Camera off confirmed by the CMO.
  refs:
  - V2 10161
- id: S5
  actor: crew
  action: The CMO restrains the patient and, with the patient's agreement, exposes the lower belly.
  success: Patient positioned.
- id: S6
  actor: robot
  rung: 2
  action: >-
    Hold the ultrasound screen where the CMO can see it and show the pre-loaded bladder scan guide (and
    the ground expert's live guidance when the link allows).
  success: Screen and guide in view.
- id: S7
  actor: crew
  action: The CMO places the probe and scans the bladder as trained.
  success: Scan done.
- id: S8
  actor: robot
  rung: 2
  action: >-
    Record the images and the device's estimate, labeled "device estimate, for human review", with the
    quality score; if the crewmember can pass some urine, record a scan before and after (both labeled
    "device estimate, for human review"). If quality is low, ask the CMO to rescan, up to <set by validation>
    times.
  success: Images and estimates recorded, or labeled non-diagnostic.
- id: S9
  actor: robot
  rung: 2
  action: >-
    Show the CMO and flight surgeon the human-entered dose log for the past <set by flight surgeon> hours,
    so the flight surgeon can review medicines that can contribute to retention (shown only; the robot
    names no medicine and gives no advice).
  success: Dose log shown, or "no dose log available".
- id: S10
  actor: robot
  rung: 2
  action: >-
    Send the images, estimate, answers and dose log to the flight surgeon at once, or queue them and show
    the CMO the onboard checklist page.
  success: Delivered or queued.
  refs:
  - V1 6003
  - V1 7001
  - V1 7003
- id: S11
  actor: ground
  action: >-
    The flight surgeon (or the CMO under the onboard checklist when there is no link) decides on catheterization,
    medicines and fluids; the robot logs the decisions.
  success: Decision logged.
  refs:
  - V1 3004
- id: S12
  actor: crew
  action: >-
    If catheterization is decided, the CMO opens the kit and does it under the onboard checklist, with
    the robot outside its keep-out distance.
  success: Procedure done and logged, or not needed.
- id: S13
  actor: robot
  rung: 2
  action: >-
    Bag waste per the biohazard procedure; clean the device and its own surfaces per RMP-INF-002; log
    supplies used and update inventory.
  success: Cleaning and inventory logged.
  refs:
  - V2 7059
- id: S14
  actor: robot
  rung: 2
  action: >-
    Set reminders, on the flight surgeon's schedule, to ask whether urine has been passed and to offer
    a repeat scan. At each scheduled time the robot prompts the crewmember on their own screen; it approaches
    only when the crewmember (or the CMO) accepts, and a prompt never starts the robot. After a catheter,
    each reminder also asks about fever, chills, or burning or cloudy urine (RF10).
  success: Reminders logged.
  refs:
  - V2 10174
decisions:
- id: D1
  after: S2
  decided_by: robot
  question: Is any red-flag answer "yes", or not answered?
  branches:
  - if: 'Yes'
    then: >-
      Apply the action listed for each matching red flag (URGENT alert, stop, or private flag) now, with
      the answer itself. For back pain with leg or saddle symptoms (RF2), open RMP-MSK-004's red-flag
      set at once; a scan may go ahead only as an adjunct, if the CMO decides, and must not be used on
      its own to decide.
  - if: Not answered
    then: Ask the CMO to ask in person; never treated as "no".
  - if: All "no"
    then: Continue to S3.
- id: D2
  after: S8
  decided_by: robot
  question: Did the images pass the device quality check within the allowed attempts?
  branches:
  - if: 'Yes'
    then: Continue.
  - if: 'No'
    then: >-
      Label "non-diagnostic (quality too low)"; report no volume; ask the ground expert to guide a rescan
      when the link allows.
- id: D3
  after: S10
  decided_by: ground
  question: Catheterize, treat or watch?
  branches:
  - if: Link live
    then: Flight surgeon decides; robot logs.
  - if: Link down or delay too long
    then: >-
      The CMO decides under the pre-approved onboard checklist; the flight surgeon reviews images and
      logs at the next contact.
- id: D4
  after: S14
  decided_by: robot
  question: >-
    Has trouble passing urine come back (no urine for the time the flight surgeon set, or symptoms return
    after a catheter is removed), or is there fever, chills, or burning or cloudy urine after a catheter?
  branches:
  - if: 'Yes'
    then: >-
      URGENT alert to the CMO and flight surgeon; restart this protocol on a crew command; for fever after
      a catheter, the CMO starts RMP-INF-001.
  - if: 'No'
    then: Keep the reminder schedule until the flight surgeon closes it.
stop_rules:
- id: X1
  trigger: The patient (or the crewmember being examined) says stop, or withdraws consent
  action: >-
    Robot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops
    unless the patient agrees to keep it running.
  level: stop
  refs:
  - V2 10165
- id: X2
  trigger: Any crewmember presses stop or takes over control
  action: Protective stop; robot holds its stopped position; only a human may restart it.
  level: safe-state
  refs:
  - V2 10165
  - V2 10168
  - V2 10174
- id: X3
  trigger: Robot fault, sensor fault, or a force, speed or reach limit is reached
  action: Protective stop and hand control to the CMO; log the fault.
  level: safe-state
  refs:
  - V2 10172
  - V2 10168
- id: X4
  trigger: CMO is asked to take over and does not respond (for example, the CMO is the patient)
  action: Robot stays stopped; the next named crewmember is called.
  level: safe-state
  refs:
  - V2 10173
  - V2 7045
- id: X5
  trigger: Loss of link, or delay longer than the limit for live guidance
  action: >-
    Only the robot pauses; crew care never waits for the link or the robot. Robot pauses any movement
    in its stopped position and tells the CMO. The CMO decides whether to continue offline under the pre-approved
    onboard checklist; the robot resumes only on the CMO's confirmation. Everything is logged and sent
    to the flight surgeon at the next contact.
  level: pause
  refs:
  - V2 10172
  - V2 10173
- id: X6
  trigger: Pain increases during the scan, or the patient asks for the robot to leave
  action: >-
    Robot backs away to its rest position and tells the CMO; the CMO continues without the robot if needed.
  level: stop
red_flags:
- id: RF1
  sign: Fever or shaking chills
  threshold: <set by clinical lead> °C or higher
  action: URGENT alert; see RMP-INF-001
  source: ng253
- id: RF2
  sign: >-
    Back pain with weakness or numbness in the legs, numbness between the legs, or loss of the feeling
    of needing to open the bowels
  threshold: any
  action: >-
    URGENT alert (possible cauda equina syndrome); open RMP-MSK-004's red-flag set at once. A bladder
    scan may go ahead only as an adjunct, if the CMO decides, and must not be used on its own to decide;
    a low volume does not rule out nerve compression.
  source: girftces
- id: RF3
  sign: Blood in the urine, or flank pain
  threshold: any
  action: URGENT alert; the flight surgeon considers RMP-RENAL-001 (the risk of stones is raised in flight)
  source: renal
- id: RF4
  sign: Unable to pass urine, with lower belly pain or urge
  threshold: any (no waiting time)
  action: URGENT alert to the CMO and flight surgeon
- id: RF5
  sign: No urine passed, without pain
  threshold: for <set by clinical lead> hours
  action: URGENT alert
- id: RF6
  sign: Device bladder-volume estimate above the alert level
  threshold: <set by clinical lead> mL (device estimate)
  action: URGENT alert to the CMO and flight surgeon
- id: RF7
  sign: The CMO cannot pass the catheter, or there is bleeding or severe pain during catheterization
  threshold: any
  action: URGENT call to the flight surgeon
- id: RF8
  sign: Retention comes back after a catheter is removed
  threshold: within <set by flight surgeon> hours
  action: URGENT alert
  source: stepaniak
- id: RF9
  sign: Confusion or feeling very unwell
  threshold: any
  action: URGENT alert
  source: ng253
- id: RF10
  sign: Fever, chills, burning or cloudy urine after catheterization
  threshold: any
  action: URGENT alert; the CMO starts RMP-INF-001
  source: ng253
safety_limits:
- id: L1
  name: Contact with the patient
  value: not allowed
  enforced: Keep-out zone in the robot controller
  refs:
  - V2 10172
- id: L2
  name: Holding or moving the ultrasound probe
  value: not allowed
  enforced: Probe is not a robot-graspable item in this protocol
- id: L3
  name: Camera recording while the patient is exposed
  value: not allowed
  enforced: Camera turned away and recording blocked until the CMO ends the exposed phase
- id: L4
  name: Keep-out distance during catheterization
  value: <set by validation>
  enforced: Distance sensing; robot holds outside it
- id: L5
  name: Robot speed near the patient
  value: <set by validation>
  enforced: Hard limit
  refs:
  - V2 10172
logs:
- id: LOG-1
  text: Every input, robot action, decision (with who made it) and hand-off, time-stamped to the second.
  refs:
  - V2 10163
  - V2 10164
- id: LOG-2
  text: Mode changes and safety-limit events are announced to the operator as they happen and logged.
  refs:
  - V2 10162
  - V2 10172
- id: LOG-3
  text: >-
    Records are stored and sent as private medical data, encrypted, with access limited to the medical
    team.
  refs:
  - V1 6003
  - V1 7003
- id: LOG-4
  text: Results are kept in a permanent, retrievable format so the flight surgeon can see trends.
  refs:
  - V1 7001
requirements:
- id: V1 3004
  use: >-
    In-mission care includes imaging, performing procedures, personal protective equipment and privacy
    considerations; administering and managing medications is a human capability.
- id: V1 3006
  use: Two crew medical officers per vehicle for NASA programs (see the commercial-flight note).
- id: V1 3007
  use: Evacuation for problems beyond onboard care; the flight surgeon decides.
- id: V1 6003
  use: Logs and alerts sent to the ground are private medical communications.
- id: V1 7001
  use: Records kept so the flight surgeon can see trends.
- id: V1 7003
  use: Secure handling of medical records.
- id: V2 7045
  use: Usable by non-physician crew, including when the CMO is the patient.
- id: V2 7046
  use: Patient, CMO, device and kits restrained.
- id: V2 7059
  use: Inventory updated with supplies used.
- id: V2 10161
  use: Robot shows its status and health.
- id: V2 10162
  use: Mode changes are announced to the operator.
- id: V2 10163
  use: Robot records and shares its data.
- id: V2 10164
  use: Shows whether the human or the robot is expected to act.
- id: V2 10165
  use: Human can override and shut down the robot.
- id: V2 10168
  use: Human can take over.
- id: V2 10172
  use: Protective stop when a limit is exceeded.
- id: V2 10173
  use: Robot stays stopped if no one takes over.
- id: V2 10174
  use: Only a human starts or restarts the robot.
validation:
- stage: simulation
  what: >-
    Scripted retention cases with red-flag answers, low-quality scans and link loss, at station, Moon
    and Mars delays.
  pass: >-
    Every red flag alerts; zero volumes reported from low-quality scans; zero cause statements; camera
    never records during the exposed phase.
- stage: training model
  what: Bladder ultrasound training model (phantom); crew scan while the robot holds the screen and records.
  pass: Image sets judged usable by an ultrasound expert at or above <set by clinical lead>.
- stage: Earth analog
  what: Analog crews with remote ultrasound experts.
  pass: Pass criteria set in advance.
- stage: flight use under flight surgeon supervision
  what: First use for scheduled scans on healthy crew.
  pass: No safety or privacy events; images usable.
signoff:
  protocol:
  - Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  - Flight surgeon, as medical authority
  - Robotics and medical-device safety board
  - Program approval for the mission
  restart_after_stop: >-
    Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot
    never restarts itself.
  each_run: >-
    Started by a crewmember. Flight surgeon agreement when the link allows; otherwise the CMO under the
    onboard checklist, with flight surgeon review of images and logs at the next contact.
change_control:
- Every change gets a new version number and a written reason; earlier versions are kept.
- >-
  Any change to steps, thresholds, limits or software that touches the patient repeats validation from
  the simulation stage before use.
- >-
  Software and learned-skill updates follow a pre-agreed change plan (what may change, how it is tested,
  how to roll back), modeled on the FDA predetermined change control plan approach.
- >-
  If a new version misbehaves, the robot rolls back to the last validated version and the event is reviewed
  by the flight surgeon and safety board.
- The robot never changes its own protocol, thresholds or limits.
comms:
  live: >-
    Low Earth orbit: the flight surgeon and a ground ultrasound expert can watch the ultrasound screen
    (not the patient) and guide the CMO live.
  loss_of_signal:
  - Only the robot pauses; crew care never waits for the link or the robot.
  - >-
    The robot pauses any movement and tells the CMO. The CMO decides whether to continue under the onboard
    checklist; the robot resumes only on the CMO's confirmation.
  - Images, estimates and answers are queued and sent at the next contact.
  mars_delay:
  - 'No live guidance: the CMO scans from the pre-loaded guide and decides under the onboard checklist.'
  - Alert levels, the catheterization plan and the reminder schedule are set before flight.
  max_delay_for_live_guidance: <set by validation>
sources:
- key: stepaniak
  text: >-
    Stepaniak PC, Ramchandani SR, Jones JA. Acute urinary retention among astronauts. Aviat Space Environ
    Med. 2007;78(4 Suppl):A5-A8.
  url: https://pubmed.ncbi.nlm.nih.gov/17511293/
  pmid: '17511293'
  supports: >-
    Case report: an astronaut who took promethazine and scopolamine before launch developed acute urinary
    retention on reaching orbit, had intermittent catheterization for the first 3 days with a single balloon-tipped
    catheter, which, because iodine solution was lacking on board and duties could not be interrupted,
    was left in place for a total of 4 days; voiding returned after day 7, retention returned on day 10
    (1 day before landing), and recurred for the first 24 hours of the next mission; obstructive, medicine-related,
    psychogenic and nerve-related causes are discussed.
- key: mtb004
  text: >-
    NASA OCHMO. Space Adaptation Sickness (SAS) and Space Readaptation Syndrome. Medical Technical Brief
    OCHMO-MTB-004 Rev A (Sept 25, 2025).
  url: https://www.nasa.gov/wp-content/uploads/2025/09/ochmo-mtb-004-space-adaptation-sickness-sas.pdf
  supports: >-
    Fluid shifts, reduced fluid intake, an altered sensation of bladder fullness, getting used to using
    the toilet in space, and changes affecting the voiding reflex can cause urinary retention in the first
    few days; crew are treated with fluid management and intermittent catheterization as needed.
- key: ermed
  text: >-
    NASA Human Research Program. Evidence Report: Risk of Adverse Health Outcomes and Decrements in Performance
    due to In-Flight Medical Conditions (2017).
  url: https://humanresearchroadmap.nasa.gov/evidence/reports/Medical.pdf
  supports: >-
    Table 2: 9 urinary retention (space adaptation) events and 9 urinary tract infections among US astronauts
    (retention: 5 female, 4 male; Shuttle, Mir and ISS through 2006); dental and urological emergencies
    have been documented among astronauts.
- key: girftces
  text: >-
    NHS England Getting It Right First Time (GIRFT). Spinal Surgery: National Suspected Cauda Equina Syndrome
    (CES) Pathway (February 2023). Also hosted by an NHS trust: https://www.gloshospitals.nhs.uk/media/documents/GIRFT_CES_Guidelines_2023_EJRRVj5.pdf
    (the copy that was read; the GIRFT-hosted copy could not be opened by automated retrieval).
  url: >-
    https://gettingitrightfirsttime.co.uk/wp-content/uploads/2023/02/National-Suspected-Cauda-Equina-Pathway-February-2023-FINAL-V1-1.pdf
  supports: >-
    Bladder scans should not be used in isolation or as a discriminator for cauda equina syndrome; if
    a patient is able to void, document the pre-void and post-void residual volumes. (The pathway's volume
    figures are not used as thresholds here.)
- key: ng253
  text: >-
    National Institute for Health and Care Excellence. Suspected sepsis in people aged 16 or over: recognition,
    assessment and early management. NICE guideline NG253 (2025, last updated Sept 23, 2026).
  url: https://www.nice.org.uk/guidance/ng253/chapter/Face-to-face-assessment
  supports: >-
    Trauma, surgery or invasive procedures in the last 6 weeks is a risk factor in the sepsis risk table;
    assess temperature, mental state and how often urine was passed.
- key: renal
  text: 'NASA Human Research Program. Evidence Report: Risk of Renal Stone Formation (2024).'
  url: https://humanresearchroadmap.nasa.gov/evidence/other/2024%20Renal%20Evidence%20Report_Final.pdf
  supports: >-
    The increased risk of renal stone formation occurs rapidly during spaceflight; stones can cause renal
    colic or severe flank pain, blood in the urine, infection or sepsis; there have been no symptomatic
    kidney stones in flight to date in US crewmembers on the ISS.
- key: tb044
  text: NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025).
  url: https://www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical-operations.pdf
  supports: >-
    The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the
    JSC 48522 medical checklist.
- key: v1
  text: 'NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023).'
  url: https://www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-rev-c-with-signature.pdf
  supports: Requirement IDs beginning V1.
- key: v2
  text: >-
    NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026),
    including Section 10.6 Automated and Robotic Systems.
  url: https://standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2
  supports: Requirement IDs beginning V2.
- key: marsdelay
  text: >-
    NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept
    Review).
  url: https://www.nasa.gov/wp-content/uploads/2024/01/mars-communications-disruption-and-delay.pdf
  supports: >-
    One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
- key: pccp
  text: >-
    U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change
    control plan for artificial intelligence-enabled device software functions (guidance).
  url: >-
    https://www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence
  supports: A model for planning, validating and rolling back software updates.
- key: faa460
  text: 'U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements.'
  url: https://www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-460
  supports: >-
    Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical
    standards for spaceflight participants.
corrections:
- date: '2026-10-20'
  version: '0.2'
  source: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 20, 2026 (batch 2 review)
  items:
  - >-
    Painful retention (unable to pass urine with lower belly pain or urge) is now URGENT with no waiting
    time; painless retention keeps the time placeholder (item 31).
  - >-
    Back pain with leg or saddle symptoms opens RMP-MSK-004's red-flag set at once; a scan is only an
    adjunct and a low volume does not rule out nerve compression (item 32).
  - >-
    Before-and-after scans are recorded when the crewmember can pass some urine; GIRFT volume figures
    are not used as thresholds (item 33).
  - >-
    Before bringing the catheter kit, the robot checks its seal and expiry and that antiseptic and spare
    catheters are in stock (item 34).
  - >-
    Added RF10: fever, chills, or burning or cloudy urine after catheterization goes URGENT to RMP-INF-001
    (item 35).
  - The Stepaniak 2007 case is described exactly as in the abstract (item 36).
  - >-
    The dose log is shown so the flight surgeon can review medicines that can contribute to retention;
    the robot names no medicine (item 37).
  - >-
    Each red-flag decision now applies the action listed for each matching red flag (URGENT alert, stop,
    or private flag), not URGENT for everything (review item 47).
  - >-
    Repeat check-ins inside an open log are now on-screen prompts; the robot approaches only when the
    crewmember or CMO accepts, and a prompt never starts it (item 49, [V2 10174]).
  - >-
    Red flags that had no source now cite one where a source fits; the rest are shown as general clinical
    red flags with no source cited (item 50).
  - >-
    The "item that failed a check" lockout is kept and now notes that it applies when RMP-PHARM-001 is
    used (item 51).
  - NASA evidence reports are cited as "approved for public release" (item 48).
- date: '2026-10-21'
  version: '0.3'
  source: Batch 5 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    Renal evidence report citation now includes "(approved for public release September 23, 2024)", matching
    RMP-RENAL-004 (batch 4 optional item 44).
  - >-
    Library-wide "safe" wording in inherited stop rules and requirement uses rewritten per batch 4 review
    item 49 (holds its stopped position / stays stopped / take over / Robot stays stopped). Schema level
    name safe-state unchanged.
- date: '2026-10-21'
  version: '0.4'
  source: Batch 6 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    Renal evidence report citation label changed from "(approved for public release September 23, 2024)"
    to "(2024)", the site's year-only citation convention.
- date: '2026-10-21'
  version: '0.5'
  source: Batch 7 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    Citation labels changed to the site's year-only convention: "(approved for public release <date>)"
    now reads "(YYYY)", and NASA-STD-3001 Volume 2 "Revision F (approved July 14, 2026)" now reads "(Rev
    F, 2026)" (sources: ermed, v2). Editorial; no change to steps, thresholds or limits.
- date: '2026-10-21'
  version: '0.6'
  source: Batch 7 review and batch 8 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    X5 and the comms loss-of-signal branch now open with "Only the robot pauses; crew care never waits
    for the link or the robot." (batch 7 review suggestion 29). Editorial; robot behaviour unchanged.