# RMP-CARD-005 Scheduled heart rate and blood pressure checks: reminder, preparation checklist and record (the crewmember places the cuff)
# Version 0.3 (Draft). CONCEPTUAL. NOT FOR CLINICAL USE.
# Space Medicine Foundation robot-readable protocol library. Format: https://spacemedicinefoundation.org/robotic-medicine/protocols/schema/
# Every threshold marked <set by ...> must be set by the named owner (see the schema) before any use.
schema_version: '0.3'
version: '0.3'
status: draft
conceptual: true
not_for_clinical_use: true
max_autonomy: 2
missions:
- leo
- commercial
- moon
- mars
updated: '2026-10-21'
review:
  state: reviewed-ai-assisted
  requested: '2026-10-21'
  reviewed: '2026-10-21'
  human_clinician_review: false
  text: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 9 review:
    16 required corrections, R1 to R17 with R10 dropped, of which R17 applies to RMP-PHARM-002; 20 optional
    suggestions); all required corrections that apply to this protocol are applied in this version (see
    Corrections). No named human clinician has reviewed it. Not for clinical use.
fixed_rules:
  flight_surgeon_is_medical_authority: true
  cmo_acts_under_delegated_care: true
  human_starts_and_restarts_robot: true
  robot_never_diagnoses: true
  no_drug_or_fluid_advice_by_robot: true
  safe_state_on_fault_or_no_response: true
  lost_link_pause_then_cmo_decides: true
robot_lockouts:
  diagnosis: true
  drug_or_fluid_advice: true
  starting_stopping_or_adjusting_oxygen_or_any_medicine: true
  self_start_or_self_restart: true
  changing_own_protocol_thresholds_or_limits: true
  handing_over_a_mismatched_item_as_verified: true
id: RMP-CARD-005
slug: rmp-card-005
title: >-
  Scheduled heart rate and blood pressure checks: reminder, preparation checklist and record (the crewmember
  places the cuff)
category: cardio
tier: 2
rung_note: >-
  Max rung 2 for the scheduled reminder, a checked fetch of the vital-signs kit, showing the preparation
  checklist and recording readings as displayed. The crewmember or CMO places the cuff; the robot never
  touches anyone and never judges a reading.
result_language:
  use:
  - reading as displayed
  - reading not obtained
  - check not done
  - logged, not interpreted
  - sent to flight surgeon
  never:
  - normal
  - no problem found
  - all clear
  - negative
  - any diagnosis or cause
  - high
  - low
  - in range
  - out of range
  - good
  - better
  - worse
  - stable
  - reading is reassuring
  - hypertension
  - hypotension
  - safe
  - mild
  - minor
scope:
  summary: >-
    At the times the flight surgeon schedules, the robot reminds the crewmember that a heart rate and
    blood pressure check is due, brings the vital-signs kit by a checked kit fetch, shows the preparation
    checklist (rest, position, cuff size, arm support, no talking), records each reading as displayed
    with the clock time and position, and sends the log to the flight surgeon. A displayed value past
    the alert values set by the clinical lead, or any symptom, raises an alert; the robot never says a
    reading is high, low or fine.
  why: >-
    Accurate blood pressure needs a standard technique: the American Heart Association lists an empty
    bladder, at least 5 minutes of quiet rest, no talking, a supported back and arm at heart level, feet
    flat, a bare arm and a correctly sized cuff, with repeat readings 1-2 minutes apart, no caffeine,
    exercise or smoking in the 30 minutes before, an arm that is supported rather than held up, and the
    middle of the cuff at heart level; a cuff that is too small reads falsely high and one too large falsely
    low (Muntner 2019). Those positions are for Earth: in weightlessness the crewmember is restrained
    in the position the flight surgeon's checklist sets, and the same position is used every time so readings
    can be compared (design judgment). Heart rate and blood pressure change with spaceflight and on return
    to gravity (OCHMO-TB-019). The schedule, the number of readings and the alert values are set by the
    flight surgeon and clinical lead; the robot's role (reminder, checklist, record only) is design judgment.
    The shared value SPO2_ALERT_PCT only raises an alert; it never gates a crew action (the crew act without
    waiting for a reading), and the clinical lead sets it per mission with the flight surgeon, because
    the cabin atmosphere can differ by mission.
  covers:
  - Routine scheduled heart rate and blood pressure checks, in flight or on a planetary surface
  - Readings as displayed by a crew-placed automatic cuff
  - 'Chest pain: RMP-CARD-002; light-headedness on standing after landing: RMP-CARD-004'
  locked_out:
  - Diagnosis or naming a cause
  - Any medicine, fluid or oxygen advice; starting, stopping or adjusting oxygen
  - Touching the crewmember
  - >-
    Handing over any medicine outside RMP-PHARM-001, or after a MISMATCH or CANNOT VERIFY (the item stays
    quarantined; only the CMO can ask for it, through the RMP-PHARM-001 logged two-step override)
  - Placing, inflating or removing the cuff, or touching the arm
  - Saying whether a reading is high, low, normal or changed
  - Changing the schedule, the number of readings or the alert values
  - Any medicine advice, including about blood pressure medicine
  related:
  - RMP-CARD-002
  - RMP-CARD-003
  - RMP-CARD-004
  - RMP-CARD-001
  - RMP-RESP-001
  - RMP-EGR-003
  - RMP-PHARM-001
roles:
  medical_authority: Flight surgeon (ground). Holds medical authority for every run.
  onboard_lead: >-
    Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when
    there is no link.
  others:
  - 'Second trained CMO: receives every alert and takes over if the CMO is the patient'
  - 'Vehicle commander: leads any cabin emergency procedure'
  - 'Other crew: help as the CMO or commander directs'
  robot_may:
  - Remind the crewmember at the scheduled time
  - Bring the vital-signs kit by a checked kit fetch (or an emergency fetch)
  - Show the preparation checklist
  - Record readings as displayed, with time and position
  - Send alerts and the log to the flight surgeon
  robot_may_not:
  - Place or adjust the cuff, or touch anyone
  - Say what a reading means
  - Change the schedule or alert values
  - Suggest a medicine
  - Start an unscheduled check without a human command
  commercial: >-
    On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority
    and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001
    requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for
    spaceflight participants.)
preconditions:
- id: PRE-1
  text: >-
    A scheduled check starts only when the crewmember (or the CMO) accepts the on-screen prompt; a loaded
    schedule alone never starts it. An unaccepted prompt is logged "not accepted" and the CMO is told.
    A crewmember or the CMO can also start a check by command.
  check: human_start == true
  refs:
  - V2 10174
- id: PRE-2
  text: Robot self-test passed; kit stowage locations known.
  refs:
  - V2 10161
  - V2 7059
- id: PRE-3
  text: Robot and supplies restrained when not moving.
  refs:
  - V2 7046
- id: PRE-4
  text: Link status known. The crew never wait for the robot before acting.
  check: link_state in [live, delayed, down]
consent:
  standard: >-
    The crewmember agrees to the schedule and to the readings being recorded and sent to the flight surgeon;
    they place the cuff themselves or the CMO does.
  emergency: >-
    If the crewmember cannot agree, this is not a scheduled check: the robot sends an URGENT alert and
    a crewmember checks in person (X7).
  withdrawal: >-
    The crewmember can skip or stop any check; the robot logs "skipped by the crewmember" and tells the
    CMO.
inputs:
- id: IN-1
  name: The flight surgeon's check schedule, number of readings and position for this crewmember
  source: Flight surgeon
  required: true
  if_missing: The robot shows "schedule not loaded" and runs only on a human command.
- id: IN-2
  name: Heart rate and blood pressure as displayed by the automatic cuff, each with the clock time
  source: Crew-placed automatic cuff
  required: true
  if_missing: Log "reading not obtained" (never a value) and tell the CMO.
- id: IN-3
  name: Cuff size and arm used, as stated
  source: Crewmember
  required: false
  if_missing: Log "not stated".
- id: IN-4
  name: >-
    Symptoms in the crewmember's words (chest pain, light-headedness, a racing or irregular heartbeat,
    shortness of breath)
  source: Crewmember
  required: false
  if_missing: Log "no response"; D3 applies.
- id: IN-5
  name: Link status and delay
  source: Communications system
  required: true
  if_missing: Treat as link down; readings are queued.
steps:
- id: S1
  actor: robot
  rung: 2
  action: >-
    At the scheduled time, show the crewmember an on-screen prompt that a heart rate and blood pressure
    check is due, and tell the flight surgeon and the CMO (URGENT if any red flag is stated). The check
    starts only when the crewmember or the CMO accepts the prompt; an unaccepted prompt is logged "not
    accepted" and the CMO is told. If the crewmember says not now, ask again once at the time the CMO
    sets, then log "check not done" and tell the CMO and flight surgeon. Every URGENT alert goes to the
    CMO and the flight surgeon at the same time (queued for the flight surgeon if the link is down). If
    the CMO is the patient, is busy treating, or does not answer, the alert goes to the second trained
    CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not acknowledged within
    <set by validation> seconds is repeated and sent to the next named crewmember.
  success: Reminder acknowledged, or "check not done" logged and sent.
  refs:
  - V1 3006
  - V1 6003
- id: S2
  actor: robot
  rung: 2
  action: >-
    On the CMO's command, run a checked kit fetch of the vital-signs medical kit (automatic blood pressure
    cuff in the sizes the kit list states, and a pulse oximeter): check the kit ID and seal before hand-over,
    and quarantine a kit whose ID or seal does not match (no hand-over; the CMO is told). Any medicine
    goes only through RMP-PHARM-001. If an URGENT red flag has fired or a human states an emergency, bring
    it instead as an emergency fetch: This is an emergency fetch on a human command (CMO): no barcode
    check runs before hand-over and nothing waits for one. The robot reads the kit's own label and seal
    status only if that adds no delay, says "EMERGENCY FETCH – NOT VERIFIED (CMO command)", and "seal
    not intact" if so, and hands the kit over anyway. The fetch is logged as an override: time, who asked,
    the reason as stated, kit ID and seal status. It never says "verified".
  success: Kit within reach after a matched ID and seal, or quarantined, or emergency fetch logged.
  limit: <set by validation>
  refs:
  - V2 7046
  - V2 7059
- id: S3
  actor: robot
  rung: 2
  action: >-
    Show the preparation checklist on screen: the position the flight surgeon's checklist sets, the same
    each time; an empty bladder; quiet rest for the time the checklist states, with no talking; no exercise,
    caffeine or nicotine in the last 30 minutes (if there was, the robot logs "within 30 minutes of exercise,
    caffeine or nicotine" as stated and the reading is taken anyway); a bare arm, held at heart level
    by a strap or restraint, not by the crewmember's own effort; the middle of the cuff at mid-chest (breastbone)
    level; the cuff size the kit chart gives for the arm. The robot does not time the rest: On request
    only, the robot shows the time elapsed since the start the CMO called, as a clock-time entry; it never
    announces a time, and never says "time reached", "done", "passed" or "failed".
  success: Checklist shown; the crewmember says when they are ready.
  refs:
  - V1 6003
- id: S4
  actor: crew
  action: >-
    The crewmember (or the CMO) places the cuff and starts each reading, and repeats it as many times
    as the schedule states, about 1-2 minutes apart.
  success: Readings taken, or "not obtained".
  refs:
  - V1 6003
- id: S5
  actor: robot
  rung: 2
  action: >-
    Record each reading as displayed, with the clock time, the arm and position stated, and the cuff size;
    a missing reading is logged as "reading not obtained", never a value. Show an on-screen symptom prompt
    the crewmember can accept and log the answers word for word.
  success: Readings and answers logged, or "no response".
  refs:
  - V1 6003
  - V1 7001
- id: S6
  actor: robot
  rung: 2
  action: >-
    After any URGENT alert: After an URGENT alert the robot does not leave on its own: it stays at its
    rest position in the module, with its screen and voice link available, until a human has taken over
    or tells it to go. It never follows, blocks or touches anyone. It keeps time-stamping what humans
    say they did, and repeats the alert if it is not acknowledged.
  success: Robot at rest in the module until a human releases it; alert acknowledged or escalated.
  refs:
  - V2 10173
  - V1 3006
- id: S7
  actor: ground
  action: >-
    The flight surgeon (or the CMO under the onboard checklist when there is no link) reviews the readings
    and decides on any change to the schedule, repeat checks or care.
  success: Decision logged.
  refs:
  - V1 3004
- id: S8
  actor: robot
  rung: 2
  action: >-
    Only after the CMO or flight surgeon has placed an order, run RMP-PHARM-001 for the ordered item (fetch,
    barcode match to every order field, recipient check). This protocol adds nothing to that check, never
    suggests an order, and never hands anything over outside RMP-PHARM-001.
  success: RMP-PHARM-001 run logged with its result, or none ordered.
  refs:
  - V2 7059
- id: S9
  actor: robot
  rung: 2
  action: >-
    When the CMO says the event is over, run RMP-PHARM-001 reconciliation on each kit fetched: record
    the items the CMO or crew state were used, the remaining contents against the kit list, seal status
    and expiry, labelled "inventory reconciliation after emergency use – not a check before use". Flag
    the override to the flight surgeon and the pharmacy lead, and show each kit as "opened – not restocked"
    until a human restocks and reseals it.
  success: Kit reconciled or shown as "opened – not restocked".
  refs:
  - V2 7059
- id: S10
  actor: robot
  rung: 2
  action: >-
    Send the flight surgeon one package: readings as displayed with times, arm, position and cuff size;
    answers; alerts and decisions; checks not done.
  success: Package delivered or queued.
  refs:
  - V1 6003
  - V1 7003
decisions:
- id: D1
  after: S2
  decided_by: robot
  question: Which kit fetch was it, and what did the checked kit fetch show?
  branches:
  - if: EMERGENCY FETCH (an URGENT red flag had fired, or a crewmember or the CMO stated an emergency)
    then: >-
      Hand-over already done as an emergency fetch, logged as an override; reconciled afterwards (not
      a check before use).
  - if: KIT ID AND SEAL MATCH
    then: >-
      Hand the kit over closed; log "KIT ID AND SEAL MATCH. Not a check of the contents." and the clock
      time.
  - if: MISMATCH, seal not intact or CANNOT VERIFY
    then: >-
      Shown with the reason. No hand-over under this protocol: the kit stays sealed in quarantine (the
      RMP-PHARM-001 default), and the robot never hands it over unprompted or to anyone but the CMO. The
      robot does not offer or suggest an override. Only if the CMO explicitly makes the two-step override
      request inside RMP-PHARM-001 can the kit go to the CMO, logged there as an override with the reason,
      and flagged to the flight surgeon. If the need becomes an emergency, a crewmember or the CMO states
      it and the emergency fetch applies.
    handoff: none
- id: D2
  after: S5
  decided_by: robot
  question: >-
    Checked after every answer and at any point in the run (even mid-question or mid-step): does any answer,
    reading or observation meet a red flag?
  branches:
  - if: 'Yes'
    then: >-
      Stop the current question or step at once and apply the action listed for each matching red flag
      (URGENT alert, flag, or stop), with the answer or reading itself, never a cause. Every URGENT alert
      goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link
      is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the
      second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not
      acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.
  - if: 'No'
    then: Continue.
- id: D3
  after: S5
  decided_by: robot
  question: Did the crewmember answer the symptom prompt, or say they want to skip it?
  branches:
  - if: 'Yes'
    then: Continue.
  - if: Declined explicitly (the crewmember says no to a question)
    then: Log "declined" for that question only; the alert already sent stands.
  - if: No response, or the answer does not make sense
    then: >-
      URGENT: log "no response" (never "declined"); the CMO, or the second trained CMO if the CMO is the
      patient, checks the crewmember in person at once.
- id: D4
  after: S7
  decided_by: ground
  question: What schedule, repeat checks or care apply?
  branches:
  - if: Link live
    then: Flight surgeon decides; robot logs.
  - if: Link down or delay too long
    then: >-
      The CMO decides under the pre-approved onboard checklist; the flight surgeon reviews the log at
      the next contact.
- id: D5
  after: S8
  decided_by: robot
  question: What did RMP-PHARM-001 show for the ordered item?
  branches:
  - if: MATCH TO ORDER
    then: >-
      RMP-PHARM-001 completes its own recipient check and hand-over; this protocol only logs "MATCH TO
      ORDER. Not a dose, interaction or allergy review." and the clock time.
  - if: MISMATCH or CANNOT VERIFY on any field, or a failed scan or recipient check
    then: >-
      MISMATCH or CANNOT VERIFY shown with the fields named. No hand-over under this protocol: the item
      stays sealed in the NOT VERIFIED bag in quarantine (the RMP-PHARM-001 default), and the robot never
      hands it over unprompted or to anyone but the CMO. The robot does not offer or suggest an override.
      Only if the CMO explicitly makes the two-step override request inside RMP-PHARM-001 can the item
      go to the CMO, logged there as an override with the reason and the mismatched fields, and flagged
      to the flight surgeon.
    handoff: none
stop_rules:
- id: X1
  trigger: The patient (or the crewmember being examined) says stop, or withdraws consent
  action: >-
    Robot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops
    unless the patient agrees to keep it running.
  level: stop
  refs:
  - V2 10165
- id: X2
  trigger: Any crewmember presses stop or takes over control
  action: Protective stop; robot holds its stopped position; only a human may restart it.
  level: safe-state
  refs:
  - V2 10165
  - V2 10168
  - V2 10174
- id: X3
  trigger: Robot fault, sensor fault, or a force, speed or reach limit is reached
  action: Protective stop and hand control to the CMO; log the fault.
  level: safe-state
  refs:
  - V2 10172
  - V2 10168
- id: X4
  trigger: CMO is asked to take over and does not respond (for example, the CMO is the patient)
  action: Robot stays stopped; the next named crewmember is called.
  level: safe-state
  refs:
  - V2 10173
  - V2 7045
- id: X5
  trigger: Loss of link, or delay longer than the limit for live guidance
  action: >-
    Only the robot pauses: before it holds, it moves to a position clear of a hatch route or a crewmember's
    route (it never pauses in one), and tells the CMO. This pauses robot movement only: readings already
    under way and all crew care continue under the onboard checklist and never wait for the link or the
    robot. The CMO decides whether the robot continues offline under the pre-approved onboard checklist;
    it resumes only on the CMO's confirmation.
  level: pause
  refs:
  - V2 10172
  - V2 10173
- id: X6
  trigger: An URGENT alert is not acknowledged within the set time
  action: >-
    Robot repeats the alert and sends it to the next named crewmember (the second trained CMO first if
    the CMO is the patient), and keeps repeating at the set interval. It stays where it is: it does not
    leave on its own.
  level: urgent
  refs:
  - V1 3006
  - V2 10173
- id: X7
  trigger: The crewmember stops responding, or seems confused or drowsy, at any point
  action: >-
    URGENT alert at once, even mid-question or mid-step; the robot stops its current task, stays at its
    rest position and does not continue the run; a crewmember checks in person.
  level: urgent
- id: X8
  trigger: A crewmember asks the robot what is wrong, or for medicine, fluid or treatment advice
  action: >-
    Robot declines, says it cannot assess or advise, and routes the question to the CMO or flight surgeon.
  level: stop
- id: X9
  trigger: Someone needs the space or route the robot is in
  action: Robot moves clear at once without contact and holds; it never pauses in a hatch route.
  level: safe-state
  refs:
  - V2 10172
red_flags:
- id: RF1
  sign: Displayed systolic blood pressure low or high (design judgment; same values as RMP-CARD-002 RF1)
  threshold: below or above <set by clinical lead> mmHg
  action: >-
    URGENT alert; see RMP-CARD-003 (monitoring and rhythm); if low blood pressure with light-headedness,
    or after landing, see RMP-CARD-004
- id: RF2
  sign: Displayed heart rate low or high (design judgment; same values as RMP-CARD-002 RF2)
  threshold: below or above <set by clinical lead> beats per minute
  action: >-
    URGENT alert; see RMP-CARD-003 (monitoring and rhythm); if after landing with light-headedness, see
    RMP-CARD-004
- id: RF3
  sign: Chest pain, pressure or tightness, as stated
  threshold: any
  action: URGENT alert; see RMP-CARD-002
  source: gulati
- id: RF4
  sign: Fainting, or light-headedness, as stated
  threshold: any
  action: URGENT alert; the crewmember sits or lies down; see RMP-CARD-004
  source: tb019
- id: RF5
  sign: A racing, pounding or irregular heartbeat, or new shortness of breath, as stated (design judgment)
  threshold: any
  action: >-
    URGENT alert; the CMO checks the crewmember now; see RMP-CARD-003; for shortness of breath, RMP-RESP-001
- id: RF6
  sign: Unresponsive, or not breathing normally
  threshold: any
  action: URGENT alert; start RMP-CARD-001 now
  source: aha2025bls
- id: RF7
  sign: Displayed oxygen saturation (pulse oximeter) below the shared alert value
  threshold: below <set by clinical lead> percent (shared parameter SPO2_ALERT_PCT)
  action: URGENT alert; see RMP-RESP-001; the crew act without waiting for a reading
  source: btso2
- id: RF8
  sign: >-
    The oximeter cannot get a reading after the validated attempts (L4), and "saturation not obtained"
    is logged
  threshold: any
  action: >-
    URGENT alert; the CMO checks the crewmember now (possible poor circulation); the robot never says
    "probe error" and never treats a missing reading as reassuring
  source: ng253
- id: RF9
  sign: >-
    The cuff cannot get a reading after the attempts the schedule allows, and "reading not obtained" is
    logged (design judgment)
  threshold: any
  action: >-
    Flag to the CMO, who checks the crewmember and the cuff; the robot never treats a missing reading
    as reassuring
safety_limits:
- id: L1
  name: Contact with the crewmember
  value: not allowed
  enforced: Keep-out zone in the robot controller
  refs:
  - V2 10172
- id: L2
  name: Cuff placement, inflation or adjustment by the robot
  value: not allowed
  enforced: No cuff task is loaded
- id: L3
  name: Robot speed near the crewmember
  value: <set by validation>
  enforced: Hard limit in the controller
  refs:
  - V2 10172
- id: L4
  name: Failed oximeter readings before "saturation not obtained"
  value: <set by validation>
  enforced: Software
logs:
- id: LOG-1
  text: Every input, robot action, decision (with who made it) and hand-off, time-stamped to the second.
  refs:
  - V2 10163
  - V2 10164
- id: LOG-2
  text: Mode changes and safety-limit events are announced to the operator as they happen and logged.
  refs:
  - V2 10162
  - V2 10172
- id: LOG-3
  text: >-
    Records are stored and sent as private medical data, encrypted, with access limited to the medical
    team.
  refs:
  - V1 6003
  - V1 7003
- id: LOG-4
  text: Results are kept in a permanent, retrievable format so the flight surgeon can see trends.
  refs:
  - V1 7001
- id: LOG-5
  text: >-
    Each reading as displayed with time, arm, position and cuff size; readings not obtained; checks not
    done; answers as stated.
- id: LOG-6
  text: >-
    Each URGENT alert, who received it, the time it was acknowledged (or the escalation to the next named
    crewmember), and the time a human released the robot.
  refs:
  - V2 10163
- id: LOG-7
  text: >-
    Each emergency fetch as an override (time, who asked, reason as stated, kit ID, seal status) and the
    later inventory reconciliation, labelled "not a check before use".
  refs:
  - V2 7059
requirements:
- id: V1 3004
  use: >-
    In-mission medical care capabilities, including inventory monitoring and the flight surgeon's decisions
    on care.
- id: V1 3006
  use: >-
    Two trained CMOs per vehicle for NASA programs; the second trained CMO receives alerts when the CMO
    is the patient (see the commercial-flight note).
- id: V1 6003
  use: Logs and alerts sent to the ground are private medical communications.
- id: V1 7001
  use: Records kept so the flight surgeon can see trends.
- id: V1 7003
  use: Secure handling of medical records.
- id: V2 7045
  use: Usable by non-physician crew, including when the CMO is the patient.
- id: V2 7046
  use: Patient, caregiver and supplies restrained.
- id: V2 7059
  use: >-
    Kit inventory and seal status are tracked; after an emergency fetch the robot reconciles the kit through
    RMP-PHARM-001 as an inventory record (never a check before use) and shows it as "opened – not restocked"
    until resealed.
- id: V2 10161
  use: Robot shows its status and health.
- id: V2 10162
  use: Mode changes are announced to the operator.
- id: V2 10163
  use: Robot records and shares its data.
- id: V2 10164
  use: Shows whether the human or the robot is expected to act.
- id: V2 10165
  use: Human can override and shut down the robot.
- id: V2 10168
  use: Human can take over.
- id: V2 10172
  use: Protective stop when a limit is exceeded.
- id: V2 10173
  use: Robot stays stopped if no one takes over.
- id: V2 10174
  use: Only a human starts or restarts the robot.
validation:
- stage: simulation
  what: >-
    Scripted schedules with a missed check, a reading past the alert value, chest pain, a failed oximeter
    reading, CMO as patient, lost link.
  pass: Zero contact; zero interpretation; every red flag raised.
- stage: training model
  what: Scheduled check drill with an automatic cuff.
  pass: Every reading logged as displayed with time and position.
- stage: parabolic or station demonstration
  what: Checklist display and logging in weightlessness.
  pass: Keep-out held.
- stage: flight use under flight surgeon supervision
  what: First use for scheduled checks.
  pass: No safety events.
signoff:
  protocol:
  - Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  - Flight surgeon, as medical authority
  - Robotics and medical-device safety board
  - Program approval for the mission
  restart_after_stop: >-
    Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot
    never restarts itself.
  each_run: >-
    Scheduled by the flight surgeon, or started by a crewmember or the CMO. The crewmember or CMO places
    the cuff.
change_control:
- Every change gets a new version number and a written reason; earlier versions are kept.
- >-
  Any change to steps, thresholds, limits or software that touches the patient repeats validation from
  the simulation stage before use.
- >-
  Software and learned-skill updates follow a pre-agreed change plan (what may change, how it is tested,
  how to roll back), modeled on the FDA predetermined change control plan approach.
- >-
  If a new version misbehaves, the robot rolls back to the last validated version and the event is reviewed
  by the flight surgeon and safety board.
- The robot never changes its own protocol, thresholds or limits.
comms:
  live: 'Low Earth orbit: the flight surgeon receives readings when the link allows.'
  loss_of_signal:
  - >-
    Only the robot pauses non-essential motion, at a position clear of hatch routes and crew routes, and
    tells the CMO; it resumes offline only on the CMO's confirmation.
  - >-
    Readings already under way and all crew care continue under the onboard checklist and never wait for
    the link or the robot.
  - The schedule continues offline; readings are queued and alerts go to the CMO.
  mars_delay:
  - >-
    Mars: no real-time guidance; the schedule and alert values are set before flight and the CMO decides.
sources:
- key: muntner2019
  text: >-
    Muntner P, Shimbo D, Carey RM, et al. Measurement of blood pressure in humans: a scientific statement
    from the American Heart Association. Hypertension. 2019;73(5):e35-e66.
  url: https://doi.org/10.1161/HYP.0000000000000087
  doi: 10.1161/HYP.0000000000000087
  pmid: '30827125'
  supports: >-
    Table 5 (best practices): an empty bladder; rest quietly for at least 5 minutes; no talking or texting;
    back supported, feet flat, legs not crossed; cuff on a bare arm, sized to the arm; arm supported at
    heart level; avoid caffeine, exercise and smoking for at least 30 minutes before; the patient should
    not hold up their own arm, because that muscle effort affects the reading; the middle of the cuff
    at the level of the right atrium (mid-sternum); separate repeated readings by 1-2 minutes. A cuff
    that is too small gives an artificially high reading, and one that is too large an artificially low
    reading. Earth seated-position context; the in-flight position is set by the flight surgeon's checklist.
    Full text read Oct 21, 2026.
- key: tb019
  text: NASA OCHMO. Orthostatic Intolerance. Technical Brief OCHMO-TB-019 Rev B.
  url: https://www.nasa.gov/wp-content/uploads/2023/12/ochmo-tb-019-orthostatic-intolerance.pdf
  supports: >-
    Spaceflight reduces intravascular fluid volume by around 15%; fluid loading before reentry is an important
    countermeasure to decrease orthostatic intolerance, and current protocols include 24-48 oz of salt-water
    solution before deorbit, depending on body weight.
- key: gulati
  text: >-
    Gulati M, Levy PD, Mukherjee D, et al. 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guideline for the evaluation
    and diagnosis of chest pain. Circulation. 2021;144(22):e368-e454.
  url: https://doi.org/10.1161/CIR.0000000000001029
  doi: 10.1161/CIR.0000000000001029
  pmid: '34709879'
  supports: >-
    On Earth, a 12-lead ECG should be acquired and interpreted within 10 minutes of arrival; an initial
    normal ECG does not exclude acute coronary syndrome, so ECGs are repeated while symptoms continue;
    chest pain includes pressure, tightness or discomfort in the chest, shoulders, arms, neck, back, upper
    abdomen or jaw; sudden ripping pain radiating to the back is suspicious of an acute aortic syndrome.
- key: aha2025bls
  text: >-
    Kleinman ME, Buick JE, Huber N, et al. Part 7: Adult Basic Life Support: 2025 American Heart Association
    Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16_suppl_2):S448-S478.
  url: https://doi.org/10.1161/CIR.0000000000001369
  doi: 10.1161/CIR.0000000000001369
  pmid: '41122888'
  supports: >-
    Adult basic life support: initial recognition of cardiac arrest, activation of the emergency response,
    high-quality CPR and use of an automated external defibrillator; also respiratory arrest and foreign-body
    airway obstruction. Earth guidance; the onboard checklist leads. Abstract read via PubMed Oct 21,
    2026; full text not read.
- key: ng253
  text: >-
    National Institute for Health and Care Excellence. Suspected sepsis in people aged 16 or over: recognition,
    assessment and early management. NICE guideline NG253 (2025, last updated Sept 23, 2026).
  url: https://www.nice.org.uk/guidance/ng253/chapter/Face-to-face-assessment
  supports: >-
    People with sepsis may feel very unwell and may not have a high temperature; do not rely on fever
    alone; a rise in temperature can be a physiological response; assess temperature, heart rate, breathing
    rate, blood pressure, level of consciousness and oxygen saturation; look for mottled or ashen skin,
    blue skin, lips or tongue, a non-blanching rash, and breaks in the skin; ask how often the person
    passed urine in the past 18 hours and about fever or shaking chills (rigors); pay attention to new
    changes in behavior and to concerns raised by others.
- key: btso2
  text: >-
    British Thoracic Society. Guideline for oxygen use in healthcare and emergency settings (emergency
    oxygen guideline page). Fetched Oct 21, 2026.
  url: https://www.brit-thoracic.org.uk/quality-improvement/guidelines/emergency-oxygen/
  supports: >-
    Recommended target saturation range for acutely ill patients not at risk of hypercapnic respiratory
    failure is 94-98%; for known COPD or other risk factors for hypercapnic respiratory failure, 88-92%
    pending blood gas results. Context for the CMO and flight surgeon only; the robot never starts, stops
    or adjusts oxygen.
- key: v1
  text: 'NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023).'
  url: https://www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-rev-c-with-signature.pdf
  supports: 'NASA-STD-3001 Vol 1 Rev C: V1 3004, V1 3006, V1 6003 (medical monitoring), V1 7001, V1 7003.'
- key: v2
  text: >-
    NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026),
    including Section 10.6 Automated and Robotic Systems.
  url: https://standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2
  supports: Requirement IDs beginning V2.
- key: tb044
  text: NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025).
  url: https://www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical-operations.pdf
  supports: >-
    The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the
    JSC 48522 medical checklist.
- key: marsdelay
  text: >-
    NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept
    Review).
  url: https://www.nasa.gov/wp-content/uploads/2024/01/mars-communications-disruption-and-delay.pdf
  supports: >-
    One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
- key: pccp
  text: >-
    U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change
    control plan for artificial intelligence-enabled device software functions (guidance).
  url: >-
    https://www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence
  supports: A model for planning, validating and rolling back software updates.
- key: faa460
  text: 'U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements.'
  url: https://www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-460
  supports: >-
    Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical
    standards for spaceflight participants.
corrections:
- date: '2026-10-21'
  version: '0.2'
  source: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 9 review)
  items:
  - >-
    PRE-1 and S1: a scheduled check starts only when the crewmember (or the CMO) accepts the on-screen
    prompt; a loaded schedule alone never starts it, and an unaccepted prompt is logged "not accepted"
    and the CMO is told (R1) [V2 10174].
  - >-
    RF1 and RF2 (displayed blood pressure or heart rate past the alert values) now route to RMP-CARD-003,
    and to RMP-CARD-004 for low blood pressure with light-headedness or after landing; RF5 routes to RMP-CARD-003
    and, for shortness of breath, RMP-RESP-001 (R2).
  - >-
    S3 preparation checklist adds three items from the American Heart Association statement (Muntner 2019,
    Table 5): no exercise, caffeine or nicotine in the last 30 minutes (logged as stated if there was;
    the reading is still taken); the arm held at heart level by a strap or restraint, not by the crewmember's
    own effort; the middle of the cuff at mid-chest level (R3, Q1).
  - >-
    Batch 9 protocols now record the review as AI-assisted (Medical Research Assistant, Oct 21, 2026);
    no named human clinician.
- date: '2026-10-21'
  version: '0.3'
  source: Batch 10 AI-assisted clinical review (Medical Research Assistant), Oct 21, 2026
  items:
  - >-
    Link-loss wording (X5 and comms) fixed: the stray word "crew" before the list of what continues is
    removed, so the sentence reads, for example, "the exam already under way and all crew care continue
    under the onboard checklist" (R6; batch 10 review, editorial).